317
General Troubleshooting
15
Problem
Possible cause
Possible solutions
The micropump issues the
“Error” signal sequence and
the diabetes manager
displays the symbol even
though the diabetes manager
and micropump are within
communication range.
The micropump turned off due to
an electronic defect (E-7).
Select the
Replace
menu and replace the reservoir. Wait at
least 30 seconds after removing the used reservoir before
connecting a new reservoir to the pump base. If the problem
persists, replace the micropump. For more information, see
chapter
9 Replacing System Components.