319
16
16
Technical Data
16.1 Micropump System
Technical data of the micropump system
Permitted insulin types
U100 insulins:
Humalog
®
, NovoLog
®
, NovoRapid
®
, Apidra
®
, Insuman
®
Infusat, Fiasp
®
Electromagnetic compatibility
The micropump system meets the EMC requirements for home healthcare environments in
accordance with IEC 60601-1-2.
Electromagnetic emission
Classified in accordance with CISPR 11, group 1, class B (residential).
Safety
The safety concept is based on a control system that consists of two microprocessors and
a supervisor microprocessor (supervising system). The control system has a dual channel
software architecture that performs all safety-relevant functions twice.
Whenever a defect or fault occurs in the control system, it is identified by the supervisor
microprocessor and vice versa. The control and supervising systems signalise errors by
means of acoustic signals and messages on the diabetes manager screen.
Communication between
micropump and diabetes manager
Bluetooth
Low Energy (BLE) wireless technology
Transmission frequency
2402–2480 MHz
man_07866364001_C_en-UK.indb 319
12.10.2018 13:10:35
Содержание Accu-Chek Solo
Страница 1: ...User s Manual Accu Chek Solo micropump system man_07866364001_C_en UK indb 1 12 10 2018 13 07 31 ...
Страница 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Страница 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Страница 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Страница 68: ...56 Putting the Diabetes Manager into Operation man_07866364001_C_en UK indb 56 12 10 2018 13 07 58 ...
Страница 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Страница 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...