348
Appendix
The calculation of the proposed correction bolus is based on your current blood
glucose result, your insulin sensitivity factor in the current time block, your target
range and on whether you plan a meal. The target BG is calculated as the mean
value from the lower and upper target limit value.
Blood glucose level
Without food intake/carbohydrates
Before a meal
Above the upper target limit value (Current BG - target BG) × insulin sensitivity
factor
(Current BG - target BG) × insulin sensitivity
meal bolus
Within target range
No correction bolus necessary.
(Current BG - target BG) × insulin sensitivity
meal bolus. The correction bolus can
be negative.
Between lower target limit value
and hypo warning limit
No bolus recommended. The correction bolus is
negative.
(Current BG - target BG) × insulin sensitivity
meal bolus. The correction bolus is
negative.
Below hypo warning limit
A hypo warning is displayed. You receive the
advice to eat fast-acting carbohydrates.
No bolus advice available.
A hypo warning is displayed. You receive the
advice to eat fast-acting carbohydrates.
No bolus advice available.
Examples of bolus advice calculations
man_07866364001_C_en-UK.indb 348
12.10.2018 13:10:40
Содержание Accu-Chek Solo
Страница 1: ...User s Manual Accu Chek Solo micropump system man_07866364001_C_en UK indb 1 12 10 2018 13 07 31 ...
Страница 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Страница 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Страница 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Страница 68: ...56 Putting the Diabetes Manager into Operation man_07866364001_C_en UK indb 56 12 10 2018 13 07 58 ...
Страница 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Страница 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...