Messages and Troubleshooting
Code
Title of maintenance
message
Possible cause/consequences
Further information
M-67
Bolus delivery failed
There is no connection to the
micropump.
Hold the diabetes manager close to the micropump.
You can deliver a quick bolus straight from the
micropump.
M-77
Operation failed
The requested operation failed.
Try again or contact your Customer Support and
Service Centre.
M-78
Outside of temperature
range
The temperature of the diabetes
manager is too high or too low.
Make sure the ambient temperature is within the
permitted range. Wait 5 minutes until the diabetes
manager has adapted itself to this temperature.
M-85
Micropump
incompatible
You tried to pair the diabetes manager
with a pump base that is incompatible.
Contact your Customer Support and Service Centre.
M-86
Micropump not started
The micropump cannot be started
because ongoing processes have not
finished yet.
Check whether you need to react to prior error
messages or maintenance messages.
Example: The prior message was
Reservoir empty
(M-21)
. Only after replacing the reservoir, will you
be able to start the micropump.
man_07866364001_C_en-UK.indb 308
12.10.2018 13:10:33
Содержание Accu-Chek Solo
Страница 1: ...User s Manual Accu Chek Solo micropump system man_07866364001_C_en UK indb 1 12 10 2018 13 07 31 ...
Страница 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Страница 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Страница 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Страница 68: ...56 Putting the Diabetes Manager into Operation man_07866364001_C_en UK indb 56 12 10 2018 13 07 58 ...
Страница 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Страница 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...