Bolus Advice
The bolus advice feature aims to bring your blood glucose level
back into the target range defined for you, by delivering an
insulin amount that was calculated in advance. However, the
bolus advice feature only makes a recommendation if you set up
bolus advice beforehand.
7.1 Overview
The bolus advice feature of the diabetes manager consists of two
components: First, a recommendation for a correction bolus to
adjust your blood glucose level if it is not within the target range.
Secondly, a recommendation for a meal bolus that covers the
carbohydrates contained in your meals.
The recommendation for the correction bolus is positive if your
current blood glucose level is above your target range. If the
blood glucose level is below the target value and a meal bolus is
recommended at the same time, a negative correction bolus
lowers the total insulin amount of the bolus advice.
If your blood glucose value is below the hypo warning limit, no
bolus advice is issued. Instead, you receive a recommendation to
consume a certain carbohydrate amount to bring your blood
glucose level back to within the target range.
Advantages of bolus advice:
j
Bolus advice is calculated based on the current blood glucose
value, your carbohydrate intake and other factors.
j
Insulin from a previous bolus that is still acting is taken into
account automatically.
j
Errors that can occur in manual bolus calculations are
prevented.
j
The blood glucose values after meals are better.
j
The metabolic adjustment is improved.
j
Therapy schemes can be adhered to more consistently.
j
Blood glucose values can be kept within the target range for a
longer time period more easily.
man_07866364001_C_en-UK.indb 129
12.10.2018 13:08:42
Содержание Accu-Chek Solo
Страница 1: ...User s Manual Accu Chek Solo micropump system man_07866364001_C_en UK indb 1 12 10 2018 13 07 31 ...
Страница 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Страница 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Страница 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Страница 68: ...56 Putting the Diabetes Manager into Operation man_07866364001_C_en UK indb 56 12 10 2018 13 07 58 ...
Страница 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Страница 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...