5 Technical data
MANUAL TME400-VCF · EN04 · 2019 March 26th
83
5.1.8.
Ground
Note
To avoid measuring errors due to electromagnetic interference, the meter
housing must be grounded with the ground connection on the lower right
section of the housing (see Figure 17: Grounding the meter).
Minimum cable cross-section:
•
length of up to 10 m:
6 mm²
•
length of 10 m or higher: 10 mm²
Figure 17: Grounding the meter
In the process, a conductive connection between the TME400 and the pipeline must
be provided as shown in the figure below.