4 Operation
MANUAL TME400-VCF · EN04 · 2019 March 26th
60
q
m min
:
minimum operating volume flow
q
m max
:
maximum operating volume flow
Example:
q
m min
= 16 m3/h
q
m max
= 250 m3/h
K
V
= 2362 pulses/m3
Hz
m
s
m
f
V
5
,
10
Impulse
2362
3600
16
3
3
min
=
=
Hz
m
s
m
f
V
164
Impulse
2362
3600
250
3
3
max
=
=
A11
Output pulse factor
The output pulse value indicates how many LF output pulses corre-
spond to one m3 (1 m³).
A12
Meter factor corrected
The meter can be adjusted by the operator, e.g. during calibration.
As a display value, this value cannot be changed. This value is only
visible if the Z27 characteristic correction is activated.
A20
Display factor
A20: Display factor for meters, including decimal places
0.01 Display with 2 decimal places
0.1 Display with 1 decimal place
1 Display without decimal places (default)
10 Display without decimal places
100 Display without decimal places
Example:
If the factor is adjusted to 0.1, the meter status is displayed with a
decimal place.
Note
If the factor is adjusted, for instance, to 10, the display value
is displayed without a decimal place.
You get the actual meter status by multiplying the display
value by 10.
This setting is marked with a "x 10" sticker (or it must be
marked).
A21
Digital output 2
mode
A21: Digital output 2 mode
0
Operating volume (default)
1
Standard volume
Содержание TME400-VC
Страница 6: ......
Страница 47: ...2 Installation MANUAL TME400 VCF EN04 2019 March 26th 41 Ex version...
Страница 48: ...2 Installation MANUAL TME400 VCF EN04 2019 March 26th 42 Ex version with current module...
Страница 49: ...2 Installation MANUAL TME400 VCF EN04 2019 March 26th 43 Non Ex version...
Страница 109: ...APPENDIX MANUAL TME400 VCF EN04 2019 March 26th 103 D Seal diagrams Will be added as soon as available...
Страница 111: ...MANUAL TME400 VCF EN04 2019 March 26th 105...