Operation
Operating Instructions for the TERZ 94 Electronic Turbine Meter
37
9. Current damping
Here you define how the current output is to be damped.
Permissible values for D are 0 to 0.9 !
Output current = (1-D) x value + D x old value
10. Correction factors for current offset and current rise
The current output is adjusted in the factory through two correction factors
(offset and rise).
In the event that a recalibration should really be necessary, there is a PC-
based program available which will define the new offset and rise values for
you.