Appendix
Manual ERZ2000-DI · EN1.2 · 2017, December 21st
117
x
When
switching on
the device
A few seconds after the device is switched on, the version numbers of BIOS and the
application appear on the bottom right of the display under the green area.
The calibration core is determined from the two first positions of the application e.g.
application = V1.0.0, then the calibration core is 1.0
x
In the
coordinate system
, in the menu
IE Software identification
, see
Figure 64:
Menu IE Software identification
.
The option can be used on the touch screen under the menu "Functions", sub-menu
"Type plate" (
chapter 2.7.4.3 Type
plate) or using the web browser.
A.3
Updating software
The update takes place by copying the new software to the SD card in the device.
Caution
x
Consult with the RMG service department to discuss how and in which
form you can obtain new software!
x
In doing so, you can ensure that archive content and device parame-
ters are not lost!
Required tools
x
Phillips screwdriver
x
SD card reader
SD card removal
x
Switch your measuring system to a safe state. If possible, discontinue the flow
through the flow computer, because there is no computing during the software up-
date and accumulating amounts are not measured.
x
Switch off the ERZ2000-DI.
x
Remove the four screws on the front panel and carefully pull it off.
Содержание ERZ2000-DI Series
Страница 6: ......
Страница 81: ...3 ELECTRICAL CONNECTIONS Manual ERZ2000 DI EN1 2 2017 December 21st 75 Figure 61 Rear wall of the ERZ2000 DI ...
Страница 99: ...4 TRANSMITTERS Manual ERZ2000 DI EN1 2 2017 December 21st 93 Figure 69 Measuring path arrangement USM GT400 ...
Страница 142: ...Certificates Manual ERZ2000 DI EN1 2 2017 December 21st 136 ...
Страница 143: ...Certificates Manual ERZ2000 DI EN1 2 2017 December 21st 137 ...
Страница 144: ...Certificates Manual ERZ2000 DI EN1 2 2017 December 21st 138 ...