ANALYTICAL COMPUTER GC 9000
.........................................................................................................................................................................................................
...............................................................................................................................................................................................................
Manual EMC 500 · EN01 · 2008-02
46
EMC Block Temperature
6
indirect
Input
and four times
right
Column X Description of coordinates
Unit
Comment(s)
1
D tB-EMC
Heading EMC block temperature
2
D tB
EMC block temperature
°C
3
D tB-in
Input value EMC block temperature
ohm
7
S tBmin
Lower fault limit, EMC block temperature
°C
8
S tBmax
Upper fault limit, EMC block temperature
°C
10
S tB-def
Default value, EMC block temperature
°C
14
S tB-af
Averaging factor (damping), EMC block temperature
21
S tB-Start
Start temp., target adjustment of EMC block temperature
°C
22
S tB-spec
Specified value, EMC block temperature
°C
23
S tB-Hyst
Hysteresis value, EMC block temperature
°C
24
S tB-Smin
Min. deviation from specified value of EMC block temperature
°C
25
S tB-Smax
Max. deviation from specified value of EMC block temperature °C
26
S tB-SCyc
Start cycles, EMC block temperature
27
S tB-Time1
Start time, EMC block temperature
Permissible values:
2 and
99 minutes (default: 30)
min 1)
28
S TB-Time2
Start time, 2nd calibration, EMC block temperature
Permissible values:
0 and
300 minutes (default: 180)
min
1) 2)
36
D tB-1108
Measured value read from EMC, block temperature TB
37
D tB-1008
Analog value read from EMC, block temperature TB
1) The changed setting will only become effective after a restart of the analytical computer
2) Attention: Input 0 is only allowed for testing operations. In normal measuring mode, start time
2 must be unequal 0!