ANALYTICAL COMPUTER GC 9000
.........................................................................................................................................................................................................
...............................................................................................................................................................................................................
Manual EMC 500 · EN01 · 2008-02
34
Current Output 4
5
indirect
Output
and three times
right
Column I
Description of coordinates
Unit
Comment(s)
1
D Current4
Heading EMC current output 4
2
D I4O
Physical value for current output 4
var
3)
3
D I4
Indication of current for output 4
mA
3)
4
C I4<
Meas. range minimum output current 4
var
1)
5
C I4>
Meas. range maximum output current 4
var
1)
6
C I4C
Calibration current output 4
mA
7
C I4O
Source for output current 4
Sup.Cal.Value / Wobbe Index / Std. Density / Rel. Density /
Inf.Cal.Value / Carbon Dioxide / Methane Number /
GC-corr. Hs / GC-corr. rn / GC-corr. CO2 / GC-corr. Ws
8
S I4-c
Correction factor for output 4
9
C I4-A
Averaging factor (damping) for output 4
var
10
C I4-mod
Mode current output 4
0-20mA / 4-20mA / Calib. Curr. / OFF
2)
11
C C4S
Source of contact output 4
Sup.Cal.Value / Wobbe Index / Std. Density / Rel. Density /
Inf.Cal.Value / Carbon Dioxide / Methane Number /
GC-corr. Hs / GC-corr. rn / GC-corr. CO2 / GC-corr. Ws /
General
12
C C4mod
Type of contact output 4
MIN cont. / MAX cont. / MIN+MAX cont. / Calib. cont. /
Test cont. / GC Offset cont.
13
C C4<
Contact range minimum, contact output 4
var
14
C C4>
Contact range maximum, contact output 4
var
1) Assigning physical limits to 0 / 4 mA or 20 mA
2) On the menu, the output current mode can be changed from 0-20 mA to 4-20 mA or calibration
current. In calibration current mode, a constant value is outputted which can be set in
coordinate I-6.
3) If a fault occurs, the current displayed will not correspond to the converted physical value.
Depending on the set mode, the current will be calculated from the default value or the last
measured value, or it will be zero.