23
Programming a new constant
You want to change the default value for the CO
2
content.
Press
9
key (Cal. Val.)
EMC Parameters
Hs 10,123
kWh/m3
Press
two times
EMC Parameters
CO2-def 1,000
Mol%
Set the switch to "Input" (in this case also the code number is sufficient)
Press the
Enter
key
The bottom line of the display turns darker and the POWER /
STANDBY LED flashes at one-second intervals to indicate the
programming mode.
Press the
"
1
"
key
EMC Parameters
CO2-def 1
Mol%
Press the
"
±,
" "
1
" "
5
"
and
"0" keys consecutively
EMC Parameters
CO2-def 1,150 Mol%
Press the
Enter
key
EMC Parameters
CO2-def
1,150
Mol%
The display turns bright
Lock the data inputted by means of the "Input"
switch
.
Programming is completed!
General information about inputting new values:
If a value is locked with the code number (user data), you must first input the correct code number
into the field Y5 in the
MODE
function. You can input values either in the short designation or
coordinate display mode. Switching over is possible at any time by pressing the
Select
key.
For values in exponential representation the "E" is entered by pressing the ± key. For this purpose it
is necessary that a comma has been entered before. So for example the input of 3E-5 is not
possible but it must be entered 3.0E-5.