ANALYTICAL COMPUTER GC 9000
.........................................................................................................................................................................................................
...............................................................................................................................................................................................................
Manual EMC 500 · EN01 · 2008-02
24
Programming a new mode
You want to change the calibration mode from "OFF" to "Calibrate".
Press the
4
key (GC Mode)
EMC-Mode
CalStart
Calibrate
Press
EMC-Mode
CalOper
OFF
Set the
switch
to "Input"
The POWER / STANDBY LED flashes at one-second intervals to indicate the programming mode,
and after you have pressed the
ENTER
key, the bottom line of the display turns darker.
Press
two times
EMC-Mode
CalOper
Calibrate
Press the
Enter
key and lock the input by means of the "input"
switch.
Programming current outputs
You can select the desired values in the columns F to I via the
5
(Output) function key and the
cursor keys. First specify the mode in field 10, then select the measured value to be outputted in
field 7 and finally program the limits in the fields 4 and 5. The constant calibration current is set in
field 6.
Example: You want to output the Wobbe index to current output 1 as current from 4 to 20 mA.
1.
Press the
5
(Output) key.
2.
Press
. (The actual mode will be displayed).
3.
Press the
Enter
key.
4.
Select the mode "4-20mA" in the filed F10 with the
and
keys.
5.
Press the
Enter
key.
6.
Press
three times.
7.
Select "Wobbeindex" in field F7 with the
and
keys.
8.
Press the
Enter
key.
Now the limits for the Wobbe index at 4 and 20 mA remain to be inputted.