Changing and calling up the programme data 

 

Parameters of the output functions 

 

16 

SYCON 
3000 H

 

6. Parameters of the output functions 

In line with the selection made in programme steps 5.1 to 5.3, supplementary data must be entered in the 
form of parameters for the outputs OUT 1 

OUT 1 

OUT 1 

OUT 1 to     OUT 3

OUT 3

OUT 3

OUT 3. 

Output function: Pulse signal 

Pulse length 

S  t  e  p    n  o  .  :                  6  .  1 
P  u  l  s  e    l  e  n  g  h  t            1  0  s 

                                   

   

   

 

 

         

             

     

The length of the pulse signal can be laid down within the range from 1 to 999 
seconds. 

 

Number of bad messages 

S  t  e  p    n  o  .  :                  6  .  2 
N  u  m  b  e  r    m  e  s  s  a  g  e  s        1  * 

                                       

   

 

 

         

#

             

       

You can programme the number of bad messages after which the output func-
tion pulse signal is activated. Values from 1 to 5 can be entered. With a value 
greater  than  1,  when  the  next  analysis  occurs  is  programmed  in  programme 
step 6.10 or 6.11. 

 

Note: 
The pulse output is always activated as soon as a bad message is displayed after an analysis and 
the  number  of  successive  bad  messages  is  the  same  as  or  greater than the value entered in pro-
gramme step 6.2 . 
In programme step 1.5, it is determined whether a bad message is to occur if the limit value is over-
run or underrun.   

 

Attention! 
If the output function "Permanent signal 1" or "Permanent signal 2" is so programmed that the ana-
lyser  performs  no  more  analyses  before  the  value  entered  in  programme  step  6.2  is  reached,  no 
further signal can be emitted. 

 

Output function: Permanent signal 1 

Activation only with a limit overrun or underrun 

 

Automatic delete function 1 

S  t  e  p    n  o  .  :                  6  .  3 
A  u  t  o  m  a  t  .  d  e  l  e  t  e    1    Y  /  N 

                                 

 

   

   

 

 

         

             

       

You  can  programme  whether  further  analyses  are  carried  out  after  the  activa-
tion  of  the  output  function  Permanent  signal  1.  If  further  analyses  are  con-
ducted, the relevant relay is automatically deactivated again if the result "Water 
good" occurs in one of the next analyses.  
If no further analyses are conducted, the relay must be deleted by hand or via 
the input "Delete relay". Moreover, an analysis must be restarted. 
Automatic deletion Yes

Yes

Yes

Yes: 

Consecutive analyses even after a bad message  

Automatic deletion No

No

No

No: 

Analysis stop in the event of a bad message 

 

Number of the bad messages 

S  t  e  p    n  o  .  :                  6  .  4 
N  u  m  b  e  r    m  e  s  s  a  g  e  s        2  * 

                                       

   

 

 

         

#

             

       

You  can  programme  the  number  of  bad  messages  (limit  overruns/underruns) 
after which the output function is activated. Values from 1 to 5 can be entered. 
With  a  value  greater  than  1,  it  is  laid  down  in  programme  step  6.9  or  6.10  
when the next analysis occurs. 

6.5

Output

Pulse lenght

6.1

s

Pulse signal

?

Autom.delete1

6.4

Output

Permanent

signal 1

6.3

5.4

no

yes

no

?

yes

Yes

No

Bad messages

6.2

*

6.5

Bad messages

*

Содержание SYCON 3000 H

Страница 1: ...OPERATING MANUAL Analytical Instrument to Measure the Total Hardness of Water SYCON 3000 H...

Страница 2: ...urement too low 6 Water hardness over underrun 6 Permanent signal 1 6 Permanent signal 2 6 Input functions 7 Start analysis 7 Stop analysis 7 Reset relay 7 Water meter 7 Output functions 8 Pulse signa...

Страница 3: ...function 2 17 Number of the bad messages 17 Activation through a fault 17 Output function Analysis is running 17 Analysis delay 17 Output function Signal relay 18 Fault messages 18 7 Activation of th...

Страница 4: ...alve is opened Each measurement begins with an adjustable flushing phase This ensures that the water from the treatment plant is measured and not the water that has remained in the supply line since t...

Страница 5: ...isplaying the last result i e Soft water or Hard water blink If both the control lamps Soft water and Hard water are blinking no analysis result is available e g after the instrument has been switched...

Страница 6: ...s arranged beneath it e g START and FLUSH Displays during the analyses B l a n c v a l u e 2 s 1 L i m i t 2 3 p p m 9 0 E N D During an analysis the phases Flush Black value Blanc value Titration and...

Страница 7: ...F O The functions of the inputs IN1 and IN2 with their current switching states are displayed A horizontal line alongside the designation means Input not active A vertical line alongside the designati...

Страница 8: ...Press the SERV key After 4 seconds the ser vicing phase is initiated You can recalibrate the electronics and switch the metering pump on and off See also the section Service setting 1 and 2 on page 2...

Страница 9: ...r sample contaminated No flush occurred No water inflow Calibration necessary Electrical defect check connector Internal measurement too high A T T E N T I O N I n t r m e a s u r t o o h i g h A n a...

Страница 10: ...analysis starts if the contact of the flow moni tor is activated longer then 10 seconds An analysis also starts if the contact is activated 5 times for 2 seconds e g if water is only taken out for a s...

Страница 11: ...6 4 the number of bad messages is programmed after which the sig nal is activated Permanent signal 2 The signal serves to trigger valves horns or signal devices that require a permanent con tact duri...

Страница 12: ...low measured value too high measured value too low These faults effect an analysis stop which must be deleted by hand Recorder output RC 2 A recorder at this output permanently registers the displaye...

Страница 13: ...seconds until this display appears C H A N G E p r o g r a m m l a n g u a g e P R O G R M L A N G U A E N D 3 Should you wish to change the language of the LCD display press the LANGUA key E N G L I...

Страница 14: ...le in step service setting 3 The correction factor needs to be determined empirically with help of a known water sample shift value positive value higher water hardness correction factor higher than 1...

Страница 15: ...analysis delay entered A new analysis is only started if the previous one has been completed For this reason an analysis interval can be greater than that programmed in this sec tion Note If a water...

Страница 16: ...the desired input function for the input INPUT 1 IN1 with the help of the key Stat Start analysis WaMe Water Meter Stop Stop analysis NoIn No Input Rest Reset relay Input IN 2 S t e p n o 3 2 S t a t...

Страница 17: ...a delay time in the range of 1 to 99 sec onds Input function Water meter Water volume 1 between the analyses S t e p n o 4 4 W a t e r v o l 1 1 0 0 c b m Enter the Water volume 1 after which a new a...

Страница 18: ...2 Output OUT 2 S t e p n o 5 2 I M P P S 1 P S 2 A N A S I G O U T 2 P e r m a n e s i g n a l 1 Select according to programme step 5 1 the desired output function for the output OUT2 Output OUT 3 S t...

Страница 19: ...1 or Permanent signal 2 is so programmed that the ana lyser performs no more analyses before the value entered in programme step 6 2 is reached no further signal can be emitted Output function Perman...

Страница 20: ...of the bad messages S t e p n o 6 6 N u m b e r m e s s a g e s 2 You can programme the number of bad messages limit overruns underruns after which the output function is activated Values from 1 to 5...

Страница 21: ...tivation of the fault relay RF Refill indicator VH Measured value too high NI No indicator Stop VL Measured value too low BV Blank value too low 7 Activation of the buzzer S t e p n o 7 1 I N B V V H...

Страница 22: ...dness of the last analysis is continuously dis played The scale of the display value can be programmed To this end a current initial value of 0 mA or 4 mA must be accorded to the hardness value of 0 d...

Страница 23: ...ote the newly entered code number in your documentation Af f f fter ter ter ter a new code number has been entered this programme step cannot be called a new code number has been entered this programm...

Страница 24: ...ion The potentiometer required for calibration is now accessible and can be actuated with the help of a screwdriver ATTENTION Replace the cap after calibration S e r v i c e s t e p n o 2 C a l i b r...

Страница 25: ...il it reaches a predetermined value W a r t u n g S c h r i t t 3 a l t 0 J N n e u 3 0 s a v e E N D E After the titration has finished the old and new determined shift value is displayed Shift the c...

Страница 26: ...eeding Feeding Feeding indicator indicator indicator indicator When commissioning the instrument for the first time the metering pump must be switched on until the indicator flows into the measuring c...

Страница 27: ...wall mounting The brackets can be mounted twisted 90 to the left or right For mounting on a plate the in strument can be mounted from the rear side without brackets Use M5 screws The depth of the thre...

Страница 28: ...OUT 1 OUT 2 Power in Protectiveearth No Nc C No Nc OUT 3 C No Nc IN 1 IN 2 RC 1 RC 2 L1 Power out n1 l1 Pow er output Pow er input Earth conductor Contact input 1 Recorder output 1 PE PE 4 AmT 500 mA...

Страница 29: ...valve with the mains output voltage Inputs The analyser is equipped with 2 identical inputs for floating switching contacts It is only through the programming in the programme steps 3 1 and 3 2 that...

Страница 30: ...ms for the regeneration of treatment plants start contact for central control valves with their own pro gramme controller barrels PLC controls stop signal contacts for osmosis plants and fault alarm s...

Страница 31: ...Programming Programming Programming Programming 1 5 MAX Activation of the horn if the water hardness is exceeded applies to all outputs 5 1 IMP Generation of a pulse signal 5 4 IMP Activation of the...

Страница 32: ...VL Activation on occurrence of the message Internal measurement too low The programming of 1 5 in the case of OUT 1 applies to all outputs Floating fault signal contact at OUT 3 Function Function Func...

Страница 33: ...Function Function The recorder serves to log the activities of the instrument Different current values are allocated to the different activities Programming Programming Programming Programming 8 1 4...

Страница 34: ...e lever and tighten the hose somewhat 3 Removing the measuring chamber a Remove the metering pump hose b Unfasten the solenoid valve connector and remove it c Loosen and remove the sensor connec tors...

Страница 35: ...e sets for maintenance Recommended spare parts for several years 90 142 Maintenance set 01 3x 2 hoses 90161 und 90163 90 362 Measuring chamber cpl without solenoid vale and without intake and outlet p...

Страница 36: ...max 3 Pulse signal Permanent signal 1 Permanent signal 2 Analysis runs Signal relay Analogue outputs 2 Output 1 Status display Analysis active Soft water Hard water Refill indicator and Measurement f...

Отзывы: