Operating displays
INFO display
4
SYCON
3000 H
INFO displays
With the help of the INFO key, different information or values can be requested. Modifications will be
described – as far as possible – in the programme section "Changing and reading out the programme
data". Only the Service telephone number can be altered during the display.
Press the KEY key repeatedly until the key function INFO is displayed for the right-hand key F3.
Press the INFO key. The first information is displayed.
Other information can be obtained by further pressing the INFO key.
9 m 5 , 3 1 c b m 2 *
L i m i t
2 . 3 m g / l
> = 9 . 0
S T A R T I N F O
Diagrams
l i m
E N D I N F O
The last 100 results of the analysis and the
programmed limit are displayed as diagrams.
So if e.g. a measurement is carried out every
10 minutes the results of the last 1000 min-
utes (aprox. 17 hours) are displayed.
Indicator type and hardness unit
I n d i c a t o r H A 1 , 2
R a n g e : 1 . 1 - 2 1 . 3
H a r d n e s s u n i t : p p m
E N D I N F O
The programmed indicator type, the associ-
ated hardness range and the unit of hardness
are displayed.
Service telephone number
S e r v i c e t e l e p h o n n o .
0 5 1 2 1 1 2 3 4 5 6 7 8
# I N F O
The Service telephone number is displayed.
Altering the telephone number:
SELECTING DIGITS:
Press the key with the "
" symbol.
CHANGING DIGITS:
Press the key with the "
#
#
#
#
" symbol.
Input states
I n p u t s t a t e s
S T A R T = - S T O P = -
E N D I N F O
The functions of the inputs IN1 and IN2 with
their current switching states are displayed.
A horizontal line "
----
" alongside the designation
means: Input not active.
A vertical line "|
||
|" alongside the designation
means: Input active.
If one of the both inputs is not in use “No In-
put” is displayed.
Output states
O u t p u t s t a t e s
P u l s e =
-
P e r m a n . 1 =
-
M e s s a g e =
-
E N D I N F O
Programmed functions of the outputs OUT1,
OUT2 and OUT3 with their current switching
states are displayed.
A horizontal line "
----
" alongside the designation
means: Output not active.
A vertical line "|
||
|" alongside the designation
means: Output active.
Analysis result counter
A n a l y s e s c o u n t e r
S O F T : 1 4 3 6
H A R D : 4 5
E N D I N F O
In the first line the number of all analyses con-
ducted with the result Water SOFT and in the
second line the number with the result Water
HARD is displayed.
The max. number totals 9'999'999. Thereafter,
the counters are reset to zero.
Software status
S o f t w a r e v e r s i o n
S Y C O N 3 0 0 0 H 0 1 0 1 . 2 5
E N D I N F O
The software is continuously serviced in the fac-
tory. Modifications are made as the need arises
to bring the product in line with new findings and
requirements.
The version number and the software imple-
mented are displayed.
Internal measurement and
blank value
I n t e r n a l M e a s u r m . 1 0 1 %
L a s t b l a n k v a l u e 9 9 %
E N D I N F O
The LED in the measurement chamber is activated
and in the first line the current value of the optical
measurement is displayed (range 0-121%).
In the second line the value of the last blank
value is displayed (range 0-121%).