Connection and programming examples
30
SYCON
3000 H
Delete key at IN 2
Function:
Function:
Function:
Function:
If a relay has been permanently activated, it can be deactivated again at the instrument or by means
of a remote switch (external delete key).
Programming:
Programming:
Programming:
Programming:
3.2 =
Rest
= Function "Reset relay/horn"
3.3 =
Rest |
= Reset of the relay when contact closed
4.3 =
1s
= Delay time for the reset of the relay = 1 second
Recorder at RC 1
Function:
Function:
Function:
Function:
The recorder serves to log the activities of the instrument. Different current values are allocated to the
different activities.
Programming:
Programming:
Programming:
Programming:
8.1 =
4 mA
=
Analysis active
8 mA
=
Analysis: soft water
12 mA =
Analysis: hard water
16 mA =
Refill indicator
20 mA =
Fault in the measurement
Recorder at RC 2
Function:
Function:
Function:
Function:
The recorder serves to log the measured values of the instrument. Initial and final values are deter-
mined in accordance with the chosen measurement range and the indicator type in programme step
1.1.
Programming:
Programming:
Programming:
Programming:
8.2 =
0 mA
=
0.0 ppm CaCO
3
20 mA =
40 ppm CaCO
3