background image

 

Messages 

 

 

 

 

SYCON 
3000 H

 

Messages 

If  the  integrated  buzzer  is  activated  in  the  course  of  a  message,  you  can  delete  it  by  pressing  the 
"HORN"  key.  The  message  in  the  LCD  display  is  only  removed  when  the  cause  of  the  message  is 
remedied or a new analysis is started manually.  
You acknowledge an activated relay by pressing the "RELAY" key. 
In  the  case  of  the  display  "Analysis  stop",  you  must  start  a  new  analysis  manually.  See  the  section 
"Manual control" on page 5. 

Refill indicator  

        I  n  d  i  c  a  t  o  r               
        r  e  f  i  l  l                     

P  o  s  s  i  b  l  e    a  n  a  l  y  s  e  s    2  4 

     

 

                   

  H  O  R  N   

Indicator  must  be refilled. On the lower right-
hand  side  the  probable  number  of  possible 
analyses is displayed. 

 

As soon as this message is displayed, a certain amount of 
indicator is available. As the indicator required per analy-
sis  is  dependent  on  the  water  hardness,  the  probable 
number  of  analyses  cannot  be  determined  exactly  in  ad-
vance.  In  calculating  the  display  value,  soft  water  with  a 
water hardness of 0.0 ppm CaCO

is assumed. 

After  each  analysis  this  value  is  recalculated  with  the  ac-
tual amount of indicator that still remains. 

 

  

If  there  is  no  indicator  when  the  instrument  is 
switched on, the display "Indicator lack Analy-
sis stop" appears immediately. 

Indicator lack analysis stop 

    A  T  T  E  N  T  I  O  N    !               
      I  n  d  i  c  a  t  o  r    l  a  c  k       
      A  n  a  l  y  s  i  s    s  t  o  p         
     

 

                      H  O  R  N   

Automatically no further analyses are started. 
Refill indicator. 

Error blank value 

    A  T  T  E  N  T  I  O  N    !               

E  r  r  o  r    b  l  a  n  k    v  a  l  u  e       

      A  n  a  l  y  s  i  s    s  t  o  p         
     

 

                      H  O  R  N   

Automatically no further analyses are started.

 

Possible causes: 
 

Instrument or sample contaminated 

 

No flush occurred 

 

No water inflow 

 

Calibration necessary 

 

Electrical defect (check connector)

 

Internal measurement too high 

    A  T  T  E  N  T  I  O  N    !               

I  n  t  r  .  m  e  a  s  u  r  .  t  o  o    h  i  g  h 

      A  n  a  l  y  s  i  s    s  t  o  p         
     

 

                      H  O  R  N   

Automatically no further analyses are started.

 

Possible causes: 
 

Indicator has not been metered out 

 

No water inflow 

 

Calibration necessary 

 

Electrical defect (check connector) 

 

Internal measurement too low 

    A  T  T  E  N  T  I  O  N    !               

I  n  t  r  .  m  e  a  s  u  r  e  .  t  o  o    l  o  w 

      A  n  a  l  y  s  i  s    s  t  o  p         
     

 

                      H  O  R  N   

Automatically no further analyses are started.

 

Possible causes:

 

 

 
 

No water inflow 

 

Calibration necessary 

 

Electrical defect (check connector)

 

Water hardness over/underrun 

    A  T  T  E  N  T  I  O  N    !               
      W a  t  e  r    h  a  r  d  n  e  s  s       
      e  x  c  e  e  d  e  d                   
     

 

                      H  O  R  N   

According  to  the  programming  in  program 
step 1.5, the overrunning or under running of 
the water hardness is displayed. 
Example: Hard water = Limit value exceeded. 

Permanent signal 1 

    A  T  T  E  N  T  I  O  N    !               

P  e  r  m  a  n  e  n  t    s  i  g  n  a  l    1     

      A  n  a  l  y  s  i  s    s  t  o  p         
     

 

                    R  E  L  A  Y   

Following an overrun or underrun of the water 
hardness (see programming step 1.5), the ac-
tivation  of  the  relay  "Permanent  signal  1"  is 
displayed. 
It  is  also  displayed  whether  an  analysis  stop 
has  occurred  or  whether  analyses  continue  to 
be conducted (see program step 6.3). 

Permanent signal 2 

    A  T  T  E  N  T  I  O  N    !               

P  e  r  m  a  n  e  n  t    s  i  g  n  a  l    2     

      A  n  a  l  y  s  i  s    s  t  o  p         
     

 

                    R  E  L  A  Y   

Following an overrun or under run of the wa-
ter  hardness  (see  program  step  1.5), the acti-
vation of the relay "Permanent signal 2" is dis-
played. This relay can be activated even in the 
case of a fault according to the programming 
in  program  step  6.7.  It  is  also  displayed 
whether  an  analysis  stop  has  occurred  or 
whether  analyses  continue  to  be  conducted 
(see program step 6.5).

 

Содержание SYCON 3000 H

Страница 1: ...OPERATING MANUAL Analytical Instrument to Measure the Total Hardness of Water SYCON 3000 H...

Страница 2: ...urement too low 6 Water hardness over underrun 6 Permanent signal 1 6 Permanent signal 2 6 Input functions 7 Start analysis 7 Stop analysis 7 Reset relay 7 Water meter 7 Output functions 8 Pulse signa...

Страница 3: ...function 2 17 Number of the bad messages 17 Activation through a fault 17 Output function Analysis is running 17 Analysis delay 17 Output function Signal relay 18 Fault messages 18 7 Activation of th...

Страница 4: ...alve is opened Each measurement begins with an adjustable flushing phase This ensures that the water from the treatment plant is measured and not the water that has remained in the supply line since t...

Страница 5: ...isplaying the last result i e Soft water or Hard water blink If both the control lamps Soft water and Hard water are blinking no analysis result is available e g after the instrument has been switched...

Страница 6: ...s arranged beneath it e g START and FLUSH Displays during the analyses B l a n c v a l u e 2 s 1 L i m i t 2 3 p p m 9 0 E N D During an analysis the phases Flush Black value Blanc value Titration and...

Страница 7: ...F O The functions of the inputs IN1 and IN2 with their current switching states are displayed A horizontal line alongside the designation means Input not active A vertical line alongside the designati...

Страница 8: ...Press the SERV key After 4 seconds the ser vicing phase is initiated You can recalibrate the electronics and switch the metering pump on and off See also the section Service setting 1 and 2 on page 2...

Страница 9: ...r sample contaminated No flush occurred No water inflow Calibration necessary Electrical defect check connector Internal measurement too high A T T E N T I O N I n t r m e a s u r t o o h i g h A n a...

Страница 10: ...analysis starts if the contact of the flow moni tor is activated longer then 10 seconds An analysis also starts if the contact is activated 5 times for 2 seconds e g if water is only taken out for a s...

Страница 11: ...6 4 the number of bad messages is programmed after which the sig nal is activated Permanent signal 2 The signal serves to trigger valves horns or signal devices that require a permanent con tact duri...

Страница 12: ...low measured value too high measured value too low These faults effect an analysis stop which must be deleted by hand Recorder output RC 2 A recorder at this output permanently registers the displaye...

Страница 13: ...seconds until this display appears C H A N G E p r o g r a m m l a n g u a g e P R O G R M L A N G U A E N D 3 Should you wish to change the language of the LCD display press the LANGUA key E N G L I...

Страница 14: ...le in step service setting 3 The correction factor needs to be determined empirically with help of a known water sample shift value positive value higher water hardness correction factor higher than 1...

Страница 15: ...analysis delay entered A new analysis is only started if the previous one has been completed For this reason an analysis interval can be greater than that programmed in this sec tion Note If a water...

Страница 16: ...the desired input function for the input INPUT 1 IN1 with the help of the key Stat Start analysis WaMe Water Meter Stop Stop analysis NoIn No Input Rest Reset relay Input IN 2 S t e p n o 3 2 S t a t...

Страница 17: ...a delay time in the range of 1 to 99 sec onds Input function Water meter Water volume 1 between the analyses S t e p n o 4 4 W a t e r v o l 1 1 0 0 c b m Enter the Water volume 1 after which a new a...

Страница 18: ...2 Output OUT 2 S t e p n o 5 2 I M P P S 1 P S 2 A N A S I G O U T 2 P e r m a n e s i g n a l 1 Select according to programme step 5 1 the desired output function for the output OUT2 Output OUT 3 S t...

Страница 19: ...1 or Permanent signal 2 is so programmed that the ana lyser performs no more analyses before the value entered in programme step 6 2 is reached no further signal can be emitted Output function Perman...

Страница 20: ...of the bad messages S t e p n o 6 6 N u m b e r m e s s a g e s 2 You can programme the number of bad messages limit overruns underruns after which the output function is activated Values from 1 to 5...

Страница 21: ...tivation of the fault relay RF Refill indicator VH Measured value too high NI No indicator Stop VL Measured value too low BV Blank value too low 7 Activation of the buzzer S t e p n o 7 1 I N B V V H...

Страница 22: ...dness of the last analysis is continuously dis played The scale of the display value can be programmed To this end a current initial value of 0 mA or 4 mA must be accorded to the hardness value of 0 d...

Страница 23: ...ote the newly entered code number in your documentation Af f f fter ter ter ter a new code number has been entered this programme step cannot be called a new code number has been entered this programm...

Страница 24: ...ion The potentiometer required for calibration is now accessible and can be actuated with the help of a screwdriver ATTENTION Replace the cap after calibration S e r v i c e s t e p n o 2 C a l i b r...

Страница 25: ...il it reaches a predetermined value W a r t u n g S c h r i t t 3 a l t 0 J N n e u 3 0 s a v e E N D E After the titration has finished the old and new determined shift value is displayed Shift the c...

Страница 26: ...eeding Feeding Feeding indicator indicator indicator indicator When commissioning the instrument for the first time the metering pump must be switched on until the indicator flows into the measuring c...

Страница 27: ...wall mounting The brackets can be mounted twisted 90 to the left or right For mounting on a plate the in strument can be mounted from the rear side without brackets Use M5 screws The depth of the thre...

Страница 28: ...OUT 1 OUT 2 Power in Protectiveearth No Nc C No Nc OUT 3 C No Nc IN 1 IN 2 RC 1 RC 2 L1 Power out n1 l1 Pow er output Pow er input Earth conductor Contact input 1 Recorder output 1 PE PE 4 AmT 500 mA...

Страница 29: ...valve with the mains output voltage Inputs The analyser is equipped with 2 identical inputs for floating switching contacts It is only through the programming in the programme steps 3 1 and 3 2 that...

Страница 30: ...ms for the regeneration of treatment plants start contact for central control valves with their own pro gramme controller barrels PLC controls stop signal contacts for osmosis plants and fault alarm s...

Страница 31: ...Programming Programming Programming Programming 1 5 MAX Activation of the horn if the water hardness is exceeded applies to all outputs 5 1 IMP Generation of a pulse signal 5 4 IMP Activation of the...

Страница 32: ...VL Activation on occurrence of the message Internal measurement too low The programming of 1 5 in the case of OUT 1 applies to all outputs Floating fault signal contact at OUT 3 Function Function Func...

Страница 33: ...Function Function The recorder serves to log the activities of the instrument Different current values are allocated to the different activities Programming Programming Programming Programming 8 1 4...

Страница 34: ...e lever and tighten the hose somewhat 3 Removing the measuring chamber a Remove the metering pump hose b Unfasten the solenoid valve connector and remove it c Loosen and remove the sensor connec tors...

Страница 35: ...e sets for maintenance Recommended spare parts for several years 90 142 Maintenance set 01 3x 2 hoses 90161 und 90163 90 362 Measuring chamber cpl without solenoid vale and without intake and outlet p...

Страница 36: ...max 3 Pulse signal Permanent signal 1 Permanent signal 2 Analysis runs Signal relay Analogue outputs 2 Output 1 Status display Analysis active Soft water Hard water Refill indicator and Measurement f...

Отзывы: