background image

 

Changing and calling up the programme data 

 

Reagents and limit value 

 

11 

SYCON 
3000 H

 

1. Reagents and limit value 

Indicator type 

 

Through  different  indicators, 
different 

measurement 

ranges  can  be  determined. 
There  are  5  indicator  types 

available.  With  the  help  of  the  "NEXT"  key,  select  the  indicator  used  in  the  in-
strument.  

 

Measurement range 

Indicator  
type 

° dH 

°E 

°F 

ppm CaCO

3

 

gpg 

mmol/Lit. 

HA 0,3 

0.02 - 0.30  0.02 - 0.37  0.03 - 0.53 

0.3 - 5.3 

0.02 - 0.31  0.0027-0.0534 

HA 1,2 

0.06 - 1.20  0.08 - 1.50  0.11 - 2.13 

1.1 - 21.3 

0.07 - 1.24  0.011 - 0.213 

HA 3 

0.15 - 3.00  0.19 - 3.75  0.27 - 5.34 

2.7 - 53.4 

0.16 - 3.12  0.027 - 0.534 

HA 30 

1.5 - 30.0  1.9 - 37.5  2.7 - 53.4 

27 - 534 

1.6 - 31.2 

0.27 - 5.34 

HA 60 

3.0 - 60.0  3.8 - 75.0  5.4 - 99.9 

54 -999 

3.2 - 62.4 

0.54 - 9.99 

 

Physical unit of water hardness 

Using the "NEXT" key, choose 
the  unit  in  which  the  water 
hardness is to be displayed: 
°dH,  °E,  °F,  ppm  CaCO

3

gpg or mmol/ltr. 

Correction factor 

The  displayed  value  is  influ-
enced  by  component  toler-
ances.  Also  the  chemical 
colour  reaction  is  influenced 

e.g.  by  non-hardness-ions.  That’s  why  it  may  be  necessary  to  adjust  the  saved 
absorption diagram to get an optimal result. So it is possible to change the zero 
point of the absorption diagram due entering a shift value (+/- 400 units) and 
to change the steepness due entering of a correction factor (0,30 – 2,00). 
It  is  also  possible  to  enter  the  shift  value automatically with help of a 0° water 
sample in step ‘service setting 3’. The correction factor needs to be determined 
empirically with help of a known water sample.  

shift value: positive value = higher water hardness  

correction factor: higher than 1,0 = higher water hardness

 

 

Limit value of the water hardness 

Determine  the  value  of  the 
limit at which a message is to 
occur  when  the  water  hard-
ness is over- or underrun (see 

programme step 1.5). 

 

Attention:

Attention:

Attention:

Attention:  The  limit  value  of  the  water  hardness is dependent on the choice of 
the  indicator  type  and  the  water  hardness.  Check  the  limit  value  of  the  water 
hardness if you have changed programme step 1.1 or 1.2. If a recorder is used, 
the scale in programme step 8.2 must also be re-assessed. 

 

Limit value monitoring 

It can be laid down whether the signalisation of the 
limit monitoring occurs in the event of an underrun 
=  soft  water  (MIN

MIN

MIN

MIN)  or  an  overrun  =  hard  water 

(MAX

MAX

MAX

MAX) of the limit value. 

 

Example: 
In monitoring an ion exchanger, the overrun is selected (MAX

(MAX

(MAX

(MAX). 

In monitoring a blending facility, the underrun  (MIN

(MIN

(MIN

(MIN) can be programmed if one wishes to monitor a minimum 

water hardness.  

S  t  e  p    n  o  .  :                  1  .  1 
T  y  p  :    H  A  1  ,  2                     

1  .  1  -  2  1  .  3    p  p  m                 

   

 

 

       

N  E  X  T 

       

       

S  t  e  p    n  o  .  :                  1  .  2 
H  a  r  d  n  e  s  s    u  n  i  t  :  p  p  m       

                                       

   

 

 

       

N  E  X  T 

       

       

S  t  e  p     n  o  .  :                  1  .  3 

0  °  p  p  m      +    5  0  F  a  c  t  .  1  ,  0  0 

     

 

         

 

       

   

   

 

 

         

             

     

S  t  e  p    n  o  .  :                  1  .  4 

L  i  m  .  v  a  l  u  e  :    9  .  0  m  g  /  l     

                         

             

   

 

 

         

             

     

S  t  e  p    n  o  .  :                  1  .  5 

L  i  m  .  v  a  l  u  e  :        M  i  n  /  M  a  x 

                                   

   

   

 

 

         

             

       

2.1

START

1.5

HA 0,3

1.1

Monitoring

Min

Max

HA 1,2

HA 3

HA 30

HA 60

I n dic a t or  t ype

1.3

Shift value

°dH

°E

°F

ppm CaCO

gpg

Water hardness

3

mmol/ltr

1.2

1.4

Limit value

Correction f actor

Содержание SYCON 3000 H

Страница 1: ...OPERATING MANUAL Analytical Instrument to Measure the Total Hardness of Water SYCON 3000 H...

Страница 2: ...urement too low 6 Water hardness over underrun 6 Permanent signal 1 6 Permanent signal 2 6 Input functions 7 Start analysis 7 Stop analysis 7 Reset relay 7 Water meter 7 Output functions 8 Pulse signa...

Страница 3: ...function 2 17 Number of the bad messages 17 Activation through a fault 17 Output function Analysis is running 17 Analysis delay 17 Output function Signal relay 18 Fault messages 18 7 Activation of th...

Страница 4: ...alve is opened Each measurement begins with an adjustable flushing phase This ensures that the water from the treatment plant is measured and not the water that has remained in the supply line since t...

Страница 5: ...isplaying the last result i e Soft water or Hard water blink If both the control lamps Soft water and Hard water are blinking no analysis result is available e g after the instrument has been switched...

Страница 6: ...s arranged beneath it e g START and FLUSH Displays during the analyses B l a n c v a l u e 2 s 1 L i m i t 2 3 p p m 9 0 E N D During an analysis the phases Flush Black value Blanc value Titration and...

Страница 7: ...F O The functions of the inputs IN1 and IN2 with their current switching states are displayed A horizontal line alongside the designation means Input not active A vertical line alongside the designati...

Страница 8: ...Press the SERV key After 4 seconds the ser vicing phase is initiated You can recalibrate the electronics and switch the metering pump on and off See also the section Service setting 1 and 2 on page 2...

Страница 9: ...r sample contaminated No flush occurred No water inflow Calibration necessary Electrical defect check connector Internal measurement too high A T T E N T I O N I n t r m e a s u r t o o h i g h A n a...

Страница 10: ...analysis starts if the contact of the flow moni tor is activated longer then 10 seconds An analysis also starts if the contact is activated 5 times for 2 seconds e g if water is only taken out for a s...

Страница 11: ...6 4 the number of bad messages is programmed after which the sig nal is activated Permanent signal 2 The signal serves to trigger valves horns or signal devices that require a permanent con tact duri...

Страница 12: ...low measured value too high measured value too low These faults effect an analysis stop which must be deleted by hand Recorder output RC 2 A recorder at this output permanently registers the displaye...

Страница 13: ...seconds until this display appears C H A N G E p r o g r a m m l a n g u a g e P R O G R M L A N G U A E N D 3 Should you wish to change the language of the LCD display press the LANGUA key E N G L I...

Страница 14: ...le in step service setting 3 The correction factor needs to be determined empirically with help of a known water sample shift value positive value higher water hardness correction factor higher than 1...

Страница 15: ...analysis delay entered A new analysis is only started if the previous one has been completed For this reason an analysis interval can be greater than that programmed in this sec tion Note If a water...

Страница 16: ...the desired input function for the input INPUT 1 IN1 with the help of the key Stat Start analysis WaMe Water Meter Stop Stop analysis NoIn No Input Rest Reset relay Input IN 2 S t e p n o 3 2 S t a t...

Страница 17: ...a delay time in the range of 1 to 99 sec onds Input function Water meter Water volume 1 between the analyses S t e p n o 4 4 W a t e r v o l 1 1 0 0 c b m Enter the Water volume 1 after which a new a...

Страница 18: ...2 Output OUT 2 S t e p n o 5 2 I M P P S 1 P S 2 A N A S I G O U T 2 P e r m a n e s i g n a l 1 Select according to programme step 5 1 the desired output function for the output OUT2 Output OUT 3 S t...

Страница 19: ...1 or Permanent signal 2 is so programmed that the ana lyser performs no more analyses before the value entered in programme step 6 2 is reached no further signal can be emitted Output function Perman...

Страница 20: ...of the bad messages S t e p n o 6 6 N u m b e r m e s s a g e s 2 You can programme the number of bad messages limit overruns underruns after which the output function is activated Values from 1 to 5...

Страница 21: ...tivation of the fault relay RF Refill indicator VH Measured value too high NI No indicator Stop VL Measured value too low BV Blank value too low 7 Activation of the buzzer S t e p n o 7 1 I N B V V H...

Страница 22: ...dness of the last analysis is continuously dis played The scale of the display value can be programmed To this end a current initial value of 0 mA or 4 mA must be accorded to the hardness value of 0 d...

Страница 23: ...ote the newly entered code number in your documentation Af f f fter ter ter ter a new code number has been entered this programme step cannot be called a new code number has been entered this programm...

Страница 24: ...ion The potentiometer required for calibration is now accessible and can be actuated with the help of a screwdriver ATTENTION Replace the cap after calibration S e r v i c e s t e p n o 2 C a l i b r...

Страница 25: ...il it reaches a predetermined value W a r t u n g S c h r i t t 3 a l t 0 J N n e u 3 0 s a v e E N D E After the titration has finished the old and new determined shift value is displayed Shift the c...

Страница 26: ...eeding Feeding Feeding indicator indicator indicator indicator When commissioning the instrument for the first time the metering pump must be switched on until the indicator flows into the measuring c...

Страница 27: ...wall mounting The brackets can be mounted twisted 90 to the left or right For mounting on a plate the in strument can be mounted from the rear side without brackets Use M5 screws The depth of the thre...

Страница 28: ...OUT 1 OUT 2 Power in Protectiveearth No Nc C No Nc OUT 3 C No Nc IN 1 IN 2 RC 1 RC 2 L1 Power out n1 l1 Pow er output Pow er input Earth conductor Contact input 1 Recorder output 1 PE PE 4 AmT 500 mA...

Страница 29: ...valve with the mains output voltage Inputs The analyser is equipped with 2 identical inputs for floating switching contacts It is only through the programming in the programme steps 3 1 and 3 2 that...

Страница 30: ...ms for the regeneration of treatment plants start contact for central control valves with their own pro gramme controller barrels PLC controls stop signal contacts for osmosis plants and fault alarm s...

Страница 31: ...Programming Programming Programming Programming 1 5 MAX Activation of the horn if the water hardness is exceeded applies to all outputs 5 1 IMP Generation of a pulse signal 5 4 IMP Activation of the...

Страница 32: ...VL Activation on occurrence of the message Internal measurement too low The programming of 1 5 in the case of OUT 1 applies to all outputs Floating fault signal contact at OUT 3 Function Function Func...

Страница 33: ...Function Function The recorder serves to log the activities of the instrument Different current values are allocated to the different activities Programming Programming Programming Programming 8 1 4...

Страница 34: ...e lever and tighten the hose somewhat 3 Removing the measuring chamber a Remove the metering pump hose b Unfasten the solenoid valve connector and remove it c Loosen and remove the sensor connec tors...

Страница 35: ...e sets for maintenance Recommended spare parts for several years 90 142 Maintenance set 01 3x 2 hoses 90161 und 90163 90 362 Measuring chamber cpl without solenoid vale and without intake and outlet p...

Страница 36: ...max 3 Pulse signal Permanent signal 1 Permanent signal 2 Analysis runs Signal relay Analogue outputs 2 Output 1 Status display Analysis active Soft water Hard water Refill indicator and Measurement f...

Отзывы: