38
•
Chapter 5: User Mode and Calibration
OX-04 Operator’s Manual
BUMP.SET (page 56)
(cont’d)
BP.EXPD
(page 60)
Defines what action must be taken if a bump test is due upon
startup.
CONFIRM
(factory setting): Press and release AIR to acknowl
-
edge that bump test is due and continue to Measuring Mode.
CANT.USE
: Cannot enter Measuring Mode until a successful
bump test is performed.
NONE
: A screen indicates that bump test is due but warmup
sequence continues.
ESCAPE
Return to the
BUMP SET
item in User Mode.
ALARM-P (page 61)
Set alarm points (WARNING, ALARM, and ALARM H) and/or reset all alarms to
their default settings.
LUNCH (page 63)
ON
: Lunch break feature is on. Instrument asks if you want to resume PEAK read
-
ings at startup.
OFF
(factory setting): Lunch break feature is off. Instrument resets PEAK readings
every time it’s turned on.
BEEP (page 63)
Set confirmation beep parameters.
BEEP.SEL
(page 64)
LED
: LED flashes and instrument vibrates based on interval
defined in
BEEP.INT
to confirm instrument is still operating.
BUZZER
: Buzzer sounds and instrument vibrates based on inter
-
val defined in
BEEP.INT
to confirm instrument is still operating.
LED+BUZ
: LED flashes, buzzer sounds, and instrument vibrates
based on interval defined in
BEEP.INT
to confirm instrument is
still operating.
BMP/CAL
: LED flashes based on interval defined in
BEEP.INT
if bump test or calibration is due regardless of whether
BP.RMDR
and/or
CAL.RMDR
are set to
ON
.
OFF
(factory setting): No alerts to confirm instrument is still oper
-
ating or that a bump test or calibration is due.
BEEP.INT
(page 64)
Confirmation alert interval. Confirmation type defined in
BEEP.SEL.
Options:
0.5
minute and
1
to
99
minutes in 1 minute increments.
The factory setting is
5
minutes.
ESCAPE
Return to the BEEP item in User Mode.
How long the back light stays on after the last button press.
Options:
0
-
255
seconds or
OFF
. The factory setting is
30
seconds.
KEY.TONE (page 65)
ON
(factory setting): Buzzer sounds when button is pressed.
OFF
: Buzzer does not sound when button is pressed.
Table 8: User Mode Menu Items
User Mode
Menu Item
Description
Содержание OX-04