60
•
Chapter 5: User Mode and Calibration
OX-04 Operator’s Manual
BP.EXPD
This item defines what indicator is given during start up when a bump test is due and
BP.RMDR
is set to
ON
(factory setting is
OFF
).
CONFIRM
(factory setting): The OX-04 gives an indicator at start up if a bump test is past due.
Press and release AIR to continue without bump testing or press and release POWER MODE to
enter User Mode and perform a bump test.
CANT.USE
: The OX-04 gives an indicator at start up that a bump test is past due. Press and
release POWER MODE to enter User Mode and perform a bump test. Pressing AIR has no effect.
A successful bump test must be performed in order to use the instrument.
NONE
: The OX-04 gives an indicator at startup that a bump test is past due. It is not necessary to
acknowledge the indicator. If desired, press POWER MODE to enter User Mode and perform a
bump test. The warm-up sequence continues on its own.
1. After entering the
BUMP.SET
menu, press AIR to scroll to
BP.EXPD
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BP.EXPD
item in the
BUMP.SET
menu.
5. See “Exiting the BUMP.SET Menu” on page 60 to return to User Mode.
Exiting the BUMP.SET Menu
1. While in the
BUMP.SET
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
BUMP.SET
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.