OX-04 Operator’s Manual
Appendix A: Maintenance Mode
•
89
Viewing the ROM/SUM (ROM/SUM)
The
ROM/SUM
screen shows the instrument’s firmware version and firmware checksum.
1. While in Maintenance Mode, press AIR to scroll to
ROM/SUM
.
2. Press and release POWER MODE. The screen shows the ROM/SUM. The ROM is the top
value and the SUM is the bottom value.
3. Press and release POWER MODE to return to the
ROM/SUM
item in Maintenance Mode.
4. See “Entering Measuring Mode (START)” on page 90 to enter Measuring Mode.
Performing a Default (M.DEF)
Performing a default operation in Maintenance Mode returns all parameters to their RKI factory
settings.
1. While in Maintenance Mode, press AIR to scroll to
M.DEF
.
2. Press and release POWER MODE.
ROM/SUM
6180
72F8
ROM/SUM
M .DEF
M .DEF
YES .MODE