Appendix A: Maintenance Mode
OX-04 Operator’s Manual
3. Use AIR to scroll to the item with the replacement date you want to view or change.
4. To change the replacement date:
a. With the desired item displayed, press and release POWER MODE.
b. Press and release POWER MODE again to set the replacement date to the current date.
5. Use the AIR button to scroll to
ESCAPE
.
6. Press and release POWER MODE to return to the
SEN.DATE
item in Maintenance Mode.
7. See “Entering Measuring Mode (START)” on page 90 to enter Measuring Mode.
Performing a Bump Test (BUMP)
See “Performing a Bump Test (BUMP)” on page 41 for a description of the
BUMP
item.
Setting Alarms to Latching or Self-Resetting (LATCH)
ON
(factory setting): The OX-04 remains in alarm until the alarm condition passes
and
POWER
MODE is pressed.
OFF
: The OX-04 automatically resets an alarm when the alarm condition passes.
1. While in Maintenance Mode, press AIR to scroll to
LATCH
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
LATCH
item in
Maintenance Mode.
5. See “Entering Measuring Mode (START)” on page 90 to enter Measuring Mode.