44
•
Chapter 5: User Mode and Calibration
OX-04 Operator’s Manual
Option C from Flow Chart
Option D from Flow Chart
• Bump test failed
•
A-CAL
set to
ON
(factory setting)
• Calibration passed
• Bump test failed
•
A-CAL
set to
ON
(factory setting)
• Calibration failed
1. A calibration immediately and
automatically starts. Continue to apply
the calibration gas.
2. The calibration time is the difference
between the
GAS.TIME
and the
CAL.TIME
values defined in the
BUMP.SET
\
SETTING
item in User
Mode.
3. The instrument shows the pass/fail
results of the bump test/calibration.
Use AIR to scroll between the results
and the readings.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE to
return to the
BUMP
screen in the
BUMP
menu.
8. Use AIR to scroll to
START
and press
and release POWER MODE to enter
Measuring Mode.
1. A calibration immediately and
automatically starts. Continue to apply
the calibration gas.
2. The calibration time is the difference
between the
GAS.TIME
and the
CAL.TIME
values defined in the
BUMP.SET
\
SETTING
item in User
Mode.
3. The instrument shows the pass/fail
results of the bump test/calibration.
The LED flashes and the buzzer
sounds. Use AIR to scroll between the
results and the readings.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE to
return to the
BUMP
screen in the
BUMP
menu.
8. Use AIR to scroll to
START
and press
and release POWER MODE to enter
Measuring Mode.
O2
0.0
CAL 60
%
O2
APLY 60
%
0.0
O2
F P
BMP / CAL
%
O2
14.5
BMP /
%
O2
1.0
/ CAL
%
AIR
AIR
O2
CAL 60
%
O2
APLY 60
%
14.5
14.5
O2
F F
BMP / CAL
%
O2
14.5
BMP /
%
O2
13.0
/ CAL
%
AIR
AIR