OX-04 Operator’s Manual
Chapter 5: User Mode and Calibration
•
43
Option A from Flow Chart
Option B from Flow Chart
• Bump test passed
• Bump test failed
• A-CAL set to OFF (factory setting is
ON)
1. The instrument indicates that the
channel passed the bump test. Use
AIR to scroll between the bump test
result and the bump test gas
reading.
2. Close the regulator.
3. Unscrew the regulator.
4. Remove the calibration cup.
5. Press and release POWER MODE
to return to the
BUMP
screen in the
BUMP
menu.
6. Use AIR to scroll to
START
and
press and release POWER MODE
to enter Measuring Mode.
1. The instrument shows if the
channel passed or failed the bump
test. The LED flashes and the
buzzer sounds. Use AIR to scroll
between the result and the reading.
2. Close the regulator.
3. Unscrew the regulator.
4. Remove the calibration cup.
5. Press and release POWER MODE
to return to the
BUMP
screen in the
BUMP
menu.
6. Use AIR to scroll to
START
and
press and release POWER MODE
to enter Measuring Mode.
7. Calibrate the OX-04 as soon as
possible.
O2
P
BMP / CAL
%
O2
1.5
BMP /
%
AIR
O2
F
BMP / CAL
%
O2
14.5
BMP /
%
AIR