56
•
Chapter 5: User Mode and Calibration
OX-04 Operator’s Manual
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
CAL.EXPD
item in the
CAL SET
menu.
5. See “Exiting the CAL SET Menu” on page 56 to return to User Mode.
Exiting the CAL SET Menu
1. While in the
CAL SET
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
CAL SET
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
Setting Bump Test Parameters (BUMP.SET)
The
BUMP.SET
menu has 5 items:
SETTING
,
BP.RMDR
,
BP.INT
,
BP.EXPD
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
BUMP.SET
.
2. Press and release POWER MODE. The
SETTING
item appears.
SETTING
The
SETTING
menu has 5 items:
GAS.TIME
,
CHECK
,
CAL.TIME
,
A-CAL
, and
ESCAPE
.
1. After entering the
BUMP.SET
menu, press AIR to scroll to
SETTING
.
2. Press and release POWER MODE. The
GAS.TIME
item appears.
BUMP .SET
SETTING
SETTING
Содержание OX-04
Страница 97: ...OX 04 Operator s Manual Warranty 97 ...