8 TROUBLESHOOTING
8.1
The possible faults with the SCENAR device and troubleshooting
methods are given in Table 2.
Table 2
Fault
Possible cause
Suggested solution
The SCENAR device
does not operate in
accordance with item
5.1.3 (for
SCENAR Home),
5.2.3 (for
SCENAR Sport)
5.3.3 (for
SCENAR Basic).
Or some adjustments
fail.
Processor mal-
function.
Reset the SCENAR device to
the default settings in accord-
ance with:
item 5.1.2 for
SCENAR Home,
item 5.2.2 for
SCENAR Sport,
item 5.3.2 for
SCENAR Basic.
If the problem persists, con-
tact the manufacturer.
The SCENAR device
emits short beeps twice
a sec when there is no
skin contact.
Low battery.
Replace the battery. If the
problem persists after the bat-
tery is replaced, contact the
manufacturer.
The SCENAR device
fails to switch ON.
The battery is
discharged.
Replace the battery.
The SCENAR device
does not emit any
sounds.
The sound is
switched OFF.
Reset the SCENAR device to
the default settings in accord-
ance with:
Item 5.1.2 for
SCENAR Home,
item 5.2.2 for
SCENAR Sport,
item 5.3.2 for
SCENAR Basic.
If the problem persists, con-
tact the manufacturer.
The SCENAR device
operates but there is no
energy on the built-in
electrode.
The protective
film is not re-
moved from the
built-in electrode.
Remove the protective film
from the built-in electrode.