Operating Manual Version 6.3-04 of 10.10.18
24
ATTENTION!
DO NOT use any power adapters to power the
SCENAR device from the line supply.
5.1.2
The SCENAR device can be reset in case of failure or when it is
required to rapidly return to the
default settings
:
stimulus energy –
minimal
;
Fm
mode –
OFF
;
Am
mode –
OFF
;
Dosing mode –
OFF
;
stimuli frequency –
90 Hz
;
sound indication –
ON
;
keyboard –
unlocked.
To reset to the default settings,
press and hold the
‘ ’
and
‘ ’
but-
tons simultaneously (for about 2 sec) until the intermittent audio signal
sounds. The reset to the default settings occurs irrespective of whether the
SCENAR device is turned on or off.
5.1.3
Repeatedly press the
‘
’
button to make sure that the
‘D’,
‘Am’, ‘Fm’
and
‘4’
LEDs light up one after another, then the ‘rolling
light’, then ‘
D
’ again and so on. The selected LED goes out after 2 seconds
if no button has been pressed.
5.1.4
Use the
‘ ’
or
‘ ’
button to adjust the stimuli energy. This is
accompanied by the LEDs flashing on indicating the energy level:
4
– from minimal energy to 25 %;
4, ‘Fm’
– from 26 to 50 %;
4, ‘Fm’, ‘Am’
– from 51 to 75 %;
4, ‘Fm’, ‘Am’, ‘D’
– from 76 to 100 %.
When the mode is preselected (the appropriate LED is glowing), press
the
‘ ’
or ‘
’
button to switch the mode on or off respectively.
5.1.5
To switch the SCENAR device off press and hold the
‘ ’
but-
ton (for about 2 sec) until the intermittent audio signal sounds.
5.1.6
The supply voltage is monitored when the SCENAR device is
on: if it is lower than (8.1 ± 0.1) V, the short audio signals sound repeated-
ly (approximately twice a second), indicating that the battery should be
replaced. Otherwise, the manufacturer shall bear no liability for incompli-
ance of the SCENAR device performance with the specifications stated in
this Operating Manual.
ATTENTION
! The low voltage sound indication is deactivated
when changing the settings or if there is a skin contact.
5.1.7
If the SCENAR device operates as described above, it is ready
for use. Otherwise, refer to Chapter 8.