100763_M
22
Rimage 2000i™ User Guide
Materials
Task
Frequency
Mandatory, continued
New fan filter
Replace the
fan filter
. Refer to the
Replace the Fan Filter
instructions on
page 24.
Monthly
Can of compressed air
Torx T10 screwdriver
Remove dust from the system. Refer
to the
Remove Dust from the Rimage
2000i
instructions on page 22.
Monthly
New ink cartridges
Replace the ink cartridges. Refer to the
Replace the Ink Cartridges
instructions
on page 25.
As needed
Optional
None needed
Clean the ink cartridges. Refer to the
Clean the Ink Cartridges
instructions
on page 24.
When print quality problems are
noticed
Torx T25 screwdriver
11/32" nut driver
5/32" Allen wrench
Align the printer. Refer to the
Align the
Printer
instructions on page 27.
As needed
Note:
The printer is aligned at the
factory.
Remove Dust from the Rimage 2000i
#
Required tools:
•
Can of compressed air
•
Torx T10 screwdriver
1. Press the
recorder buttons
to open the recorder trays.
2. Power off the
Rimage 2000i
.
3. Disconnect the
cables
from the back of the system.
n
n
Caution:n
Some cables in this system are latched.
Equipment damage may occur if cables are not
disconnected appropriately.
•
Power cord
•
USB cable
Back Panel
USB Cable
Power Cord
E
ng
lis
h
E
ng
lis
h