6 Operations and Functions
6-4. Other
functions
- 30 -
6-4. Other functions
<Suppression Function>
The sensors used with the detector are influenced by environmental changes (temperature, humidity and
other characteristics) in no small measure, which affects the reading. Therefore, the reading might be
fluctuated even in a normal environment.
This function obscures influences by environmental changes or other factors that have no meaning for your
management rules of alarm. This function is used to hide (suppress) the fluctuation of the reading under the
setting value, indicating 20.9 vol%.
NOTE
The suppression function is factory-set. The standard setting value is ±0.5 vol% (for 20.9 vol%).
In the maintenance mode, this function is disabled and the fluctuation of the reading under the setting
value is displayed.
<<Example>>
Suppression Disabled
A fluctuation around zero is displayed as the
reading.
Suppression Enabled
A fluctuation under the setting value is hidden with 20.9
vol%.
(Zero suppression)
CAUTION
A reading under zero is suppressed with the 10% FS suppression.
A reading that gets 10% FS or more under zero is displayed as "-0.0", which prevents an accurate
gas detection and needs the zero adjustment.