7 Maintenance
7-3. Calibration
method
- 40 -
7-3. Calibration method
Perform a calibration in each mode (zero
adjustment mode and span adjustment mode)
using the calibration gas.
Zero adjustment gas (collected in a gas sampling
bag)
Calibration gas (collected in a gas
sampling bag)
Gas sampling bags
<Zero Adjustment "2-1">
This is used to perform the zero adjustment.
When the detector is used to prevent oxygen deficiency, zero adjustment is not required.
NOTE
Before starting the zero adjustment, provide the detector with the zero adjustment gas and wait until
the indicator is stabilized.
2-1
Press SET key.
Current Concentration Value Display
Press the SET key to perform the zero
adjustment.
Under Zero Adjustment (CAL. is displayed)
Wait for a while until the adjustment is
completed.
Zero Adjustment Completed
It automatically returns to 2-1 after PASS is
displayed.
* If the zero adjustment failed, it automatically
returns to 2-1 after FAIL is displayed.
Return to 2-1.
Return to 2-1.
<SD-1OX>
Calibration adapter
Gas
sampling
bag
Exhaust air
-> OUT
IN
↑
Flow meter
Needle valve
Pump
P