7 Maintenance
7-2. Regular
maintenance mode
- 37 -
<Fault Alarm Test "2-0.2">
<LED Test "2-0.3">
2-0.2
Press SET key.
Fault Test ON/OFF
Select either ON/OFF.
Switch ON and press the SET key to trigger
the fault alarm.
Return to OFF and press the SET key to
cancel the test.
(Pressing the MENU/ESC key also enables
to cancel the test and to go back to the
original state.)
2-0.3
Press SET key.
LED Test ON/OFF
Select either ON/OFF.
Switch ON and press the SET key to turn on
the LEDs (POWER, ALM, FAULT) and to
display the gas concentration of 8.8.8.8.
Return to OFF and press the SET key to
cancel the test.
(Pressing the MENU/ESC key also enables
to cancel the test and to go back to the
original state.)
WARNING
Be careful to perform the test because the contact and gas concentration output can be activated by
a fault alarm test even in the maintenance mode depending on setting.