Chapter 5 Functions of the Front Panel of VSA
5-40 RSA3000E User Guide
Marker->
Sets the other system parameters (such as center frequency and reference level) by
using the current marker values. Press any key under Marker-> to activate a
marker automatically if no marker is currently enabled.
Mkr->CF
Sets the center frequency of the analyzer to the frequency of the current marker.
If Position marker is selected, the center frequency will be set to the frequency
of the current marker.
If Delta marker is selected, the center frequency will be set to the frequency of
the Delta marker.
Mkr->CF Step
Sets the center frequency step of the analyzer to the frequency of the current
marker.
If Position marker is selected, the center frequency step will be set to the
frequency of the current marker.
If Delta marker is selected, the center frequency step will be set to the
frequency difference between the Delta marker and the reference marker.
Mkr->Start
Sets the start frequency of the analyzer to the frequency of the current marker.
If Position marker is selected, the start frequency will be set to the frequency of
the current marker.
If Delta marker is selected, the start frequency will be set to the frequency of the
Delta marker.
Mkr->Stop
Sets the stop frequency of the analyzer to the frequency of the current marker.
If Position marker is selected, the stop frequency will be set to the frequency of
the current marker.
If Delta marker is selected, the stop frequency will be set to the frequency of the
Delta marker.