Meters Control (Only Available for EMI Mode)
Sets the parameters of the meter.
1.
Meters
1)
Select Meter
Specifies the currently selected meter to "Meter1", "Meter2", or "Meter3".
2)
Meter
Enables or disables the selected meter.
When enabled, the histogram of the selected meter is displayed in the
meter display area of the user interface, and the corresponding
detector type is selected to perform the measurement.
When disabled, the selected meter is not displayed and the system will
not perform the measurement.
3)
Detector
Sets the detector of the selected meter, including "Pos Peak", "Quasi Peak",
"CISPR Average", "RMS Average", "Average(Vol)", and "Neg Peak".
Note: "Quasi Peak", "CISPR Average", "RMS Average", and "Average(Vol)"
are mutually exclusive. You can at most select two among the three items
("Quasi Peak", "CISPR Average", and "RMS Average").
4)
Limit
Sets the limits of the selected meter.
5)
Limit State
Enables or disables the limit line of the selected meter.
2.
Dwell Time
Sets the dwell time of the meter detector.
3.
Peak Hold Time
1)
Peak Hold Type
Sets the max hold time type of the meter to "Adjust" or "Infinite".
When "Infinite" is selected, the peak hold line of the selected meter will
not be reset. The Adjust Time menu is grayed out and disabled.
When "Adjust" is selected, the peak hold line of the selected meter will
be reset to the current signal value after the set peak hold time. At this
time, you can use the Adjust Time menu to set the peak hold time.
2)
Adjust Time
Sets the peak hold time of the meter.
4.
Reset Peak Hold
Resets the peak hold lines of all the currently enabled meters to the current