Chapter 4 Functions of the Front Panel of EMI
4-28 RSA3000E User Guide
(Re)measure
Sets the type of the final measurement to "Curr Signal", "All Signals", or "Marked
Sigs".
Remarks:
If you select the type of the marked signal, but the marked signal is not found in
the signal table, then re-measurement cannot be executed.
If a scan sequence is currently executed, then the Re(measure) menu is
grayed out and disabled.
Detectors
Sets the detector parameters related to the final measurement. If a scan sequence is
currently executed, then the Detectors menu is grayed out and disabled.
Note: when these detector parameters are set, the detector parameters under the
Marker Func key will also be changed accordingly.
1.
Select Detector
Select the detector to be "Detector 1", "Detector 2", or "Detector 3".
2.
Dwell Time
Sets the dwell time of the selected detector.
3.
Detector
Sets the type of the selected detector. The available detector types include Pos
Peak, Quasi Peak, CISPR Average, RMS Average, Average, and Neg Peak. You
can also turn off the detector.
4.
Limit for Delta
Sets the limit line of the current detector measurement, so as to obtain the limit
delta data in the signal table. The available limits include "Limit 1" through "Limit
6".
Auto Couple
When you enable "Auto Couple" function, all the manual/auto settings in the current
measurement mode will be set to "Auto". This operation does not affect other
measurement modes.
In auto state, the auto coupled parameters are changed with their coupled
parameters. The auto coupling operation will ensure the optimal performance of the
instrument. After the operation, all the auto coupled parameters will immediately be
automatically reset based on the coupled parameters.