Chapter 4 Functions of the Front Panel of EMI
4-22 RSA3000E User Guide
5.
Test Trace
Sets the trace for the current limit line test.
6.
Delete Limit
Deletes the current limit line. The data of the current limit line will be cleared
and they will be restored to factory defaults.
7.
Delete All Limits
Deletes all limit lines. After you press the menu, the data of all the limit lines will
be cleared and they will be restored to factory defaults.
Signal Table
Sets the parameters for the signal table.
The signal table is displayed below the user interface, as shown in the figure below.
Figure 4-1 Signal Table
Parameter
Description
NO.
indicates signal sequence number.
Checkbox
"
√
" indicates that the signal is marked.
Trc
indicates the measured trace.
Freq
indicates the frequency of the searched frequency point that
meets the peak condition.
Peak Amp
indicates positive peak detector amplitude.
Displays the corresponding amplitude when the current trace
detector is pos peak after the search operationis executed.
Displays the corresponding detection amplitude of the final
measurement detector 1 type after the final measurement is
performed.
If there is no measurement data, "--" is displayed.
QP Amp
indicates quasi peak detector amplitude
Displays the corresponding amplitude when the current trace
detector is quasi-peak after the search operationis executed.
Displays the corresponding detection amplitude of the final
measurement detector 2 type after the final measurement is
performed.
If there is no measurement data, "--" is displayed.