RIGOL
User’s Guide for DSA1000 Series
2-43
2.3.4
Source (For DSA1030-TG)
2.3.4.1
Source
When the Source is On, a signal with the same frequency of the current sweep signal
will be output from the GEN OUTPUT 50Ω terminal on the front panel (You are
recommended to use the tracking signal ranges from 10 MHz to 3 GHz). The Power of
the signal could be set through the menu key. Two working modes are available in the
source. They are Power Sweep output and Fixed Power output (when Power Sweep is
Off).
The Source function is valid only when the instrument is DSA1030-TG.
2.3.4.2
TG level
Sets the output power of the signal from the tracking generator (TG). In Power Sweep
mode, this parameter indicates the start power of the sweep.
You can use the numeric keys, knob or direction keys to modify this parameter. For
more details, please refer to “Parameter Input”.
Table 2-24 TG level
Parameter
Explanation
Default
0 dBm
Range
-20 dBm to 0 dBm
Unit
dBm, -dBm, mV, uV
Knob Step
1 dBm
Direction Key Step 10 dBm
2.3.4.3
TG Level Offset
Sets the offset of the output power from the TG when gains or losses occur between
the TG output and peripheral equipment, in order to display the actual power value.
Содержание DSA1030
Страница 1: ...RIGOL User s Guide DSA1000 Series Spectrum Analyzer Sept 2014 RIGOL Technologies Inc...
Страница 2: ......
Страница 22: ......
Страница 179: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 3 Figure 3 1 Measuring a Sine signal...
Страница 181: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 5 Figure 3 2 Use the frequency counter to measure...
Страница 186: ...RIGOL Measurement Examples 3 10 User s Guide for DSA1000 Series Figure 3 4 Phase noise measurement...
Страница 188: ...RIGOL Measurement Examples 3 12 User s Guide for DSA1000 Series Figure 3 5 N dB bandwidth measurement result...
Страница 190: ...RIGOL Measurement Examples 3 14 User s Guide for DSA1000 Series Figure 3 6 Time domain power measurement...
Страница 192: ...RIGOL Measurement Examples 3 16 User s Guide for DSA1000 Series Figure 3 7 Adjacent channel power measurement result...
Страница 194: ...RIGOL Measurement Examples 3 18 User s Guide for DSA1000 Series Figure 3 8 Channel power measurement result...
Страница 196: ...RIGOL Measurement Examples 3 20 User s Guide for DSA1000 Series Figure 3 9 Occupied bandwidth measurement result...
Страница 198: ...RIGOL Measurement Examples 3 22 User s Guide for DSA1000 Series Figure 3 10 Emission bandwidth measurement result...
Страница 200: ...RIGOL Measurement Examples 3 24 User s Guide for DSA1000 Series Figure 3 11 C N Ratio measurement result...
Страница 204: ...RIGOL Measurement Examples 3 28 User s Guide for DSA1000 Series Figure 3 13 Harmonic Distortion measurement result...
Страница 211: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 35 Figure 3 17 Amplitude correction...
Страница 213: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 37 Figure 3 18 The frequency response of a filter...
Страница 214: ......