Method 2: (Use Harmonic Distortion function to measure)
Procedures:
1.
Device attachment
Connect the output terminal of the generator to the RF INPUT 50Ω terminal on
the front panel of the analyzer.
2.
Use Harmonic Distortion function to measure
(1)
Reset the analyzer
-
Press Preset
(2)
Set the parameters
-
Press FREQ
-
Press Center Freq and enter 100 MHz.
-
Press Meas
-
Press Meas Fctn and enter the lower menu.
-
Page down
-
Press Harmo Dist
(3)
Modify the harmonic distortion settings
-
Press Meas Setup
-
Press NO.of Harmo and enter 3.
3.
Measurement result
The measurement results shown in the lower window are:
Fundamental wave: 100 MHz, -9.90 dBm
Second harmonic: 200 MHz, -41.76 dBm
Third harmonic: 300 MHz, -50.98 dBm
Total harmonic distortion (THD): 2.72%
See figure below.
Содержание DSA1030
Страница 1: ...RIGOL User s Guide DSA1000 Series Spectrum Analyzer Sept 2014 RIGOL Technologies Inc...
Страница 2: ......
Страница 22: ......
Страница 179: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 3 Figure 3 1 Measuring a Sine signal...
Страница 181: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 5 Figure 3 2 Use the frequency counter to measure...
Страница 186: ...RIGOL Measurement Examples 3 10 User s Guide for DSA1000 Series Figure 3 4 Phase noise measurement...
Страница 188: ...RIGOL Measurement Examples 3 12 User s Guide for DSA1000 Series Figure 3 5 N dB bandwidth measurement result...
Страница 190: ...RIGOL Measurement Examples 3 14 User s Guide for DSA1000 Series Figure 3 6 Time domain power measurement...
Страница 192: ...RIGOL Measurement Examples 3 16 User s Guide for DSA1000 Series Figure 3 7 Adjacent channel power measurement result...
Страница 194: ...RIGOL Measurement Examples 3 18 User s Guide for DSA1000 Series Figure 3 8 Channel power measurement result...
Страница 196: ...RIGOL Measurement Examples 3 20 User s Guide for DSA1000 Series Figure 3 9 Occupied bandwidth measurement result...
Страница 198: ...RIGOL Measurement Examples 3 22 User s Guide for DSA1000 Series Figure 3 10 Emission bandwidth measurement result...
Страница 200: ...RIGOL Measurement Examples 3 24 User s Guide for DSA1000 Series Figure 3 11 C N Ratio measurement result...
Страница 204: ...RIGOL Measurement Examples 3 28 User s Guide for DSA1000 Series Figure 3 13 Harmonic Distortion measurement result...
Страница 211: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 35 Figure 3 17 Amplitude correction...
Страница 213: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 37 Figure 3 18 The frequency response of a filter...
Страница 214: ......