RIGOL
User’s Guide for DSA1000 Series
2-89
The difference between the Cont peak and the Signal Track
The Cont peak focuses on the searching for the maximum in the current channel,
while in Signal Track, the system will search for and mark the point (with no more
than 3 dB variation in amplitude) near the marker before Signal Track was enabled
as well as set the frequency of this point as the center frequency.
2.5.4.7
Search Para
Defines the conditions of peak search for various peak searching. A real peak should
meet the value of both “PK Excursn” and “PK Thresh”.
1.
PK Excursn
Sets the excursion between the peak and the minimum amplitude on both sides
of it. Peaks whose excursions are beyond the specified excursion are treated as
real peaks.
Table 2-59 PK Excursn
Parameter
Explanation
Default
10 dB
Range
0 dB to 200 dB
Unit
dB
Knob Step
1 dB
Direction Key Step 1 dB
2.
PK Thresh
Assigns a minimum for the peak amplitude. A real peak should always be beyond
the limit you specified.
Table 2-60 PK Thresh
Parameter
Explanation
Default
-90 dBm
Range
-200 dBm to 0 dBm
Unit
dBm
、
-dBm
、
mV
、
uV
Knob Step
1 dBm
Direction Key Step 1 dBm
Содержание DSA1030
Страница 1: ...RIGOL User s Guide DSA1000 Series Spectrum Analyzer Sept 2014 RIGOL Technologies Inc...
Страница 2: ......
Страница 22: ......
Страница 179: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 3 Figure 3 1 Measuring a Sine signal...
Страница 181: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 5 Figure 3 2 Use the frequency counter to measure...
Страница 186: ...RIGOL Measurement Examples 3 10 User s Guide for DSA1000 Series Figure 3 4 Phase noise measurement...
Страница 188: ...RIGOL Measurement Examples 3 12 User s Guide for DSA1000 Series Figure 3 5 N dB bandwidth measurement result...
Страница 190: ...RIGOL Measurement Examples 3 14 User s Guide for DSA1000 Series Figure 3 6 Time domain power measurement...
Страница 192: ...RIGOL Measurement Examples 3 16 User s Guide for DSA1000 Series Figure 3 7 Adjacent channel power measurement result...
Страница 194: ...RIGOL Measurement Examples 3 18 User s Guide for DSA1000 Series Figure 3 8 Channel power measurement result...
Страница 196: ...RIGOL Measurement Examples 3 20 User s Guide for DSA1000 Series Figure 3 9 Occupied bandwidth measurement result...
Страница 198: ...RIGOL Measurement Examples 3 22 User s Guide for DSA1000 Series Figure 3 10 Emission bandwidth measurement result...
Страница 200: ...RIGOL Measurement Examples 3 24 User s Guide for DSA1000 Series Figure 3 11 C N Ratio measurement result...
Страница 204: ...RIGOL Measurement Examples 3 28 User s Guide for DSA1000 Series Figure 3 13 Harmonic Distortion measurement result...
Страница 211: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 35 Figure 3 17 Amplitude correction...
Страница 213: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 37 Figure 3 18 The frequency response of a filter...
Страница 214: ......