DG4000 User's Guide
8-7
unit from the pop-up unit menu or use the direction keys and knob to
modify the current value.
Press
Type to set the statistic type of the statistic parameter.
Cyclic: restart the statistic operation when the number of statistic
operations reaches the specified statistic length.
Stop: stop the statistic operation when the number of statistic operations
reaches the specified statistic length.
Note: When the statistic function is turned on,
Statist Para is grayed out and
disabled.
3.
Select the measurement mode
Press
Speed to set the measurement mode of the counter to "Full" or
"Normal".
Full: the counter performs measurements exactly according to the specified
gate time.
Normal: the counter cannot perform measurements exactly according to the
specified gate time due to the effect of the other tasks of the system. Some
Measurement values might be lost if the signal under test varies quickly.
4.
Select the display mode
After the statistic function is turned on, press
Display to set the display mode
of the statistic results to "Digital" or "Curve", as shown in Figure 8-2 and
Figure 8-3.
Figure 8-2 Digital Display Interface of the Statistic Results
Содержание DG4062
Страница 1: ...RIGOL User s Guide DG4000 Series Function Arbitrary Waveform Generator Dec 2015 RIGOL TECHNOLOGIES INC...
Страница 2: ......
Страница 14: ...RIGOL XII DG4000 User s Guide...
Страница 31: ...Chapter 1 Quick Start RIGOL DG4000 User s Guide 1 5 116 7 153 160 7 104 9 12 100 Figure 1 4 Side View...
Страница 92: ......
Страница 128: ......
Страница 140: ......
Страница 150: ......
Страница 210: ......
Страница 213: ...Chapter 11 Remote Control RIGOL DG4000 User s Guide 11 3 3 4 5 6...
Страница 218: ......
Страница 222: ......