DG4000 User's Guide
8-4
measurement.
Press
HFSuppre to enable or disable the high-frequency suppression
function.
Note: Enable the high-frequency suppression function when low-frequency
signal with lower than 1kHz frequency is measured to filter out the
high-frequency noise interference; disable the high-frequency suppression
function when high-frequency signal with greater than 1kHz frequency is
measured.
7.
Gate Time
Press
Gate Time to set the gate time of the measurement system and the
default is "1 ms".
GateT1
1 ms
GateT2
10 ms
GateT3
100 ms
GateT4
1 s
GateT5
10 s
GateT6
>10 s
8.
Select Meas
Press
Select Meas to select the type of parameter to be measured by the
counter. The counter can measure the frequency ("Freq"), period ("Period"),
duty cycle ("Duty"), positive pulse width ("+Width") and negative pulse width
("–Width")
and the default is "Freq".
9.
Statist Set
Press
Statist Set to set the statistic function of the counter. For the detailed
information, please refer to "
10.
Auto
Press
Auto and the instrument will select a proper gate time automatically
according to the characteristics of the signal under test.
Содержание DG4062
Страница 1: ...RIGOL User s Guide DG4000 Series Function Arbitrary Waveform Generator Dec 2015 RIGOL TECHNOLOGIES INC...
Страница 2: ......
Страница 14: ...RIGOL XII DG4000 User s Guide...
Страница 31: ...Chapter 1 Quick Start RIGOL DG4000 User s Guide 1 5 116 7 153 160 7 104 9 12 100 Figure 1 4 Side View...
Страница 92: ......
Страница 128: ......
Страница 140: ......
Страница 150: ......
Страница 210: ......
Страница 213: ...Chapter 11 Remote Control RIGOL DG4000 User s Guide 11 3 3 4 5 6...
Страница 218: ......
Страница 222: ......