background image

 

36 | 

Page              Rigel Medical 

SP-SIM 

user Manual V2.2 

 

 

4.6.  Memory options 

 
 
This feature is used to view information about the Rigel SP-SIM memory status. Press 

 (F4), select setup using the up/down keys and select Memory Options by pressing 

the press 

 (F4) to accept.  

 

 

 
The display shows the number of asset records currently stored, the remaining memory 
space, number of assets deleted and number of assets in upload memory. 
 
The nature of Flash memory is such that when records are deleted only the reference to 
the  data  is  removed.  The  data  still  remains  and  will  use  memory  space.  The  memory 
must be erased in order to release memory used by deleted records.  
 
If the Erase (F2) key is pressed, a prompt is shown below. To erase the memory press 

button (F4). 

 

 

 
 

Содержание SP-SIM

Страница 1: ...Innovating Together RIGEL SP SIM SP02 simulator Copyright 2016 SEAWARD GROUP Last Update 20th April 2016 Instruction Manual 370A559 Revision 2 2...

Страница 2: ...Any items returned for calibration outside of the 13 month period stated above may not be eligible for the second 12 month section of warranty The second 12 month section of the warranty begins at th...

Страница 3: ...t from the SEAWARD GROUP This also applies to accompanying drawings and diagrams Due to a policy of continuous development the SEAWARD GROUP reserves the right to alter the equipment specification and...

Страница 4: ...the appropriate local separate collection system for electrical and electronic products or contact your local supplier for further information Please dispose of this product according to local regulat...

Страница 5: ...listed we declare under our sole responsibility that the product Rigel SP SIM SPO2 simulator Conforms with the relevant Directives and conformity is indicated by the symbol i e Conformit Europ enne S...

Страница 6: ...R curves 8 2 2 Getting started 9 2 3 Performing SP02 simulations 10 2 4 Unique use of ICONS 11 3 Logging on 12 4 Setting up your tester 13 4 1 Test Sequences 13 4 2 Asset Trace Variables 25 4 3 Syste...

Страница 7: ...C 56 8 3 Configuration Data 57 8 4 Clone Data 60 8 5 Load TnT Logo 61 9 About 63 10 Trouble shooting 64 11 Maintaining the Rigel SP SIM 65 11 1 Cleaning the SP SIM 65 11 2 User Maintenance 65 11 3 Ret...

Страница 8: ...ctions must be followed to avoid danger to persons Important follow the documentation This symbol indicates that the operating instructions must be adhered to in order to avoid danger If the SP SIM is...

Страница 9: ...test results for record keeping details of up to10 000 devices can be stored Manufacturer specific test programs can be uploaded via the Bluetooth technology while probe and monitor are tested at the...

Страница 10: ...y data capturing and asset management Intuitive User Interface Fast and easy navigation using graphics display Graphics Display Allowing real time display of simulation data Bluetooth Communication PC...

Страница 11: ...phone number with every item you Test n Tag Automatic barcode generation to provide easy use of the barcode scanner thus speeding up your test time The unique Test n Tag label provides test status and...

Страница 12: ...uracy of simulation when used with the corresponding R curves Resolution Range Repeatability 1 steps 30 60 5 1 steps 60 90 3 1 steps 90 100 1 Based on using the same probe and monitor setup Note that...

Страница 13: ...laying the settings from the previous simulation for all parameters and can be repeated by simply pressing the green start key For both the Rigel UNI and SP SIM the user will need to select SPO2 Simul...

Страница 14: ...the top side of the PULS R Parameters can be varied to validate the monitor and probe including Technology E g PULS R Masimo Nellcor SPO2 simulated Heart Rate Supported Default curves Beijing Choice...

Страница 15: ...not only provides highly visible and easy to follow menu structures but also allows the user to operate the tester using intuitive icons to speed up their test routines Below are of some of the icons...

Страница 16: ...ovided and select the new user In case a password is set for the new user please enter the password case sensitive In case a password is lost and the unit is blocked please call our helpdesk on 44 0 1...

Страница 17: ...or create test sequences see 4 1 Asset Trace Variables Generate default list of variables see 4 2 Systems Config Configure default test options see 4 3 Blue Tooth Favourites Setup your Blue Tooth dev...

Страница 18: ...ngineering units e g Joules mmHg PSI CmH2O etc are available 4 1 2 View Delete or Copy an existing Test Routine Preset test sequences cannot be deleted or changed by the user however alterations are p...

Страница 19: ...ing test sequences Print Test Sequence will provide an overview of the Test Sequence on the Favourite Blue Tooth Printer See 4 4 for help on setting up Blue Tooth devices Delete Sequence will remove t...

Страница 20: ...se or use language specific characters press the button F1 and select the required entry To Delete a character press the button F2 To return to the previous menu without changes press the button F3 Co...

Страница 21: ...s shown above The SPO2 sub menu will allow you to set the SPO2 simulation mode This is only required once in the whole sequence Open the drop down box using the arrow keys and select the required SPO2...

Страница 22: ...to navigate and key pad to enter the required data When the right settings have been achieved press the button F4 to save the changes or press F3 to return to the previous menu without changes 4 1 4 I...

Страница 23: ...ailable tests using the up and down keys Once the desired test has been highlighted confirm using the OK button F4 and finally confirm to insert into the test Sequence Pressing the button F3 will retu...

Страница 24: ...SP SIM must be set to the start settings Start Chronometer Test the next settings Chronometer Test and when the final settings have been inserted a number of next settings can be inserted the SP SIM m...

Страница 25: ...version certain damage or upgrades To create customised inspections use the insert Custom Test function using the instructions below and set the engineering units to blank Hint Create a unique range o...

Страница 26: ...will define the nature of the test eg test Instruction prior to leak test Use the drop down box to choose a preset instruction and engineering unit or simply type in the boxes provided Newly entered...

Страница 27: ...gel SP SIM will return to the previous meny without changes being made 4 1 7 Create a new Test Sequence To create a new Test Sequence enter the Menu button F4 from the home screen select Setup followe...

Страница 28: ...t 1 has been created as Test Name To change to upper or lower case or use language specific charaters press the button F1 and select the required entry To Delete a character press the button F2 To ret...

Страница 29: ...a new test needs inserting using the up and down keys Note Tests will be inserted before the highlighted position not after To modify a test sequence please refer to 4 1 3 4 2 Asset Trace Variables Th...

Страница 30: ...ing from or adding to the list of default items Each variable entered during the automatic test will automatically be added to the list for future use Trace Variables that are de activated set to NO w...

Страница 31: ...se the left and right keys to select from the following New Test Automatically brings up the next test screen Download Automatically down loads the test results to the PC see 8 1 Print Label Automatic...

Страница 32: ...ct your local Rigel sales office UK 44 0 191 5878701 4 3 5 Date Time Allows the user to set the current date time and preferred formats Press the Date Time button F2 The following menu will be display...

Страница 33: ...of which can store up to 3 accessories of the same type The three sub lists are shown below Barcode Scanner Printer Computer To select Bluetooth Favourites press the button F4 followed by Setup in the...

Страница 34: ...the search Note If more than 10 Bluetooth devices have been found the list will show the last 10 devices found In the following example the search has identified Bluetooth devices named SELLT023 SELP...

Страница 35: ...red for the device to be added to the Favourites list Note Seaward and Rigel accessories do not require a PIN For other Bluetooth devices refer to the manufacturer s documentation Note Devices that ar...

Страница 36: ...hanges and return to the previous screen SELLT023 and Elite 36 5011 are now a Bluetooth Favourite and the Rigel SP SIM will automatically establish a connection with SELLT023 when using functions that...

Страница 37: ...ter the User Admin feature press the button F1 from the home screen select Setup followed be User Admin The following menu will appear Navigate through this menu using the Up Down Keys and select by p...

Страница 38: ...Always On Power Save Back light on for 30 secs after each key press Use the up down left and right keys to select User Rights For information only To change the user rights see 4 5 4 The F2 button al...

Страница 39: ...password and repeat the steps above See menu below for an overview of the Change Password screen 4 5 4 User Profile This feature allows the administrator to setup new and existing users and inhibit o...

Страница 40: ...ows the number of asset records currently stored the remaining memory space number of assets deleted and number of assets in upload memory The nature of Flash memory is such that when records are dele...

Страница 41: ...the Setup menu will default to Factory Settings including All Asset Trace Variables Test Sequences User Admin Test Codes and System Config Warning Restore Factory Settings cannot be undone and will re...

Страница 42: ...example to test a selection of SPO2 value Heart rate Manufacturer R curves Skin colour Perfusion Arrhythmia s upgrade Artefacts upgrade Chronometer function Checking of LED quality both Red and InfraR...

Страница 43: ...a is under constant development Please monitor our website for new firmware releases New firmware can be uploaded by the end user and does not void the calibration of the device Please go to http www...

Страница 44: ...n colours Perfusion Select the required signal strength to simulate different patient conditions ie infant obese average Enter a value between 0 20 in steps of 0 1 where 5 is average perfusion Once th...

Страница 45: ...taken from one simulation setting to the next Note The SP SIM simulates both Red and Infrared data through either of the Red or InfraRed LED s at any given time This allows the user to test the quali...

Страница 46: ...the Asset ID number 25 character max and date time of test Multiple entries can be applied using the same Asset ID number In this case the memory will hold all entries and differentiate using the test...

Страница 47: ...n Trace variables can be added once the above data has been entered and confirmed See next chapters for more details 6 1 Performing an Automatic Test The automatic mode provides an option to test all...

Страница 48: ...ee 3 2 3 Use the up down left and right keys to select the required Test Sequence 4 Enter the required test period in months 5 Returning to the Asset Details screen the button F2 provides the user wit...

Страница 49: ...l not appear during the automatic tests Note that a maximum of 40 individual entries are available for each Trace Variable Additional entries will overwrite on a first in first out basis When all info...

Страница 50: ...in this example is a chronometer test measuring the time taken for the monitor to establish a reading from 90 to 85 target SPO2 After 6 seconds the SPO2 monitor displayed the required value at which t...

Страница 51: ...s Depending on the settings in Menu Setup System Config see 4 3 2 a number of After Test options are possible New Test Automatically brings up the next test screen No further options or viewing of res...

Страница 52: ...uetooth Favorites See 4 4 A detailed application note is available from the Rigel website on the used of Test n Tag printers with the Rigel SP SIM Please visit http www rigelmedical com knowledgebase...

Страница 53: ...bel will print a pass fail label on the thermal printer 50 mm roll Print Result will print the result on the thermal printer 50 mm roll 6 4 View Results Press the View result Button to review the data...

Страница 54: ...menu The following screen provides an overview of all available records in the memory identified by Asset ID Site and Pass or Fail Use the up and down arrow keys to scroll through the database When th...

Страница 55: ...st Details screen press the button F4 to view the following options View Results will display the results of the test see section 7 1 View Sequence will display the test Sequence User Comment will all...

Страница 56: ...Test Details screen select button F4 Use the up and down keys to select the view results option Press the button F4 to display the view text results screen see below Use the button F2 to return to ma...

Страница 57: ...eria fill in the required fields by either using the keyboard or drop down boxes Note The entered data must be identical to the data stored with the test This includes visible and invisible SPACE char...

Страница 58: ...int All Labels Will download all records within the search criteria to the Test n Tag printer setup in Bluetooth favourites Download to PC Will download all records within the search criteria to the P...

Страница 59: ...sing the button F4 8 1 Download to PC This feature is used to download records to a PC Select Download to PC and press the button F4 to accept The Rigel SP SIM will now attempt to make a Bluetooth con...

Страница 60: ...er data The No of assets field shows the number of records that will be downloaded Ensure that the PC application is ready to receive data then press the Send button F4 A progress bar as shown below w...

Страница 61: ...from our knowledge base Please visit www rigelmedical com knowledgebase 8 3 Configuration Data This feature can be used to transfer store or modify the current Trace Variables User Profiles and Comme...

Страница 62: ...ng used with the P C The baud rate will default to the correct value of 57600 In the application select Start Capture from the Capture menu in the task bar Enter a valid file name for the text file co...

Страница 63: ...convenience Note The file transfer can be made using any suitable PC application such as the Seaward Bluetooth Downloader Data Transfer exe or Hyperterminal capable of sending and receiving a text fi...

Страница 64: ...fer is not successful check that the format of the configuration file exactly matches that specified Note The Config file must be ASCII text and the last entry must be the END statement 8 4 Clone Data...

Страница 65: ...T Logo A bitmap logo can be transferred to the Rigel SP SIM for use with the Seaward Test n Tag print system The logo must be in BMP format with a maximum size of 320 pixels by 240 pixels Press button...

Страница 66: ...window allowing you to select the Bitmap file Select the required Bitmap and click on Open The Bluetooth Downloader will now transfer the file to the Rigel SP SIM When the transfer is complete the Ri...

Страница 67: ...t Calibration Date Serial Number Blue Tooth ID Select ABOUT from the menu below Select using the F4 key and use the scoll keys F3 to toggle between the different information screens Ensure you have th...

Страница 68: ...in an over lit area ie cover the sensor if required 4 Remove the power supply from the SP SIM The SP SIM unit can sometimes give an incorrect or noisy simulation when used with certain Pulse Oximeter...

Страница 69: ...is a rugged quality instrument However care should always be taken when using transporting and storing this type of equipment Failure to treat the product with care will reduce both the life of the i...

Страница 70: ...739 Fax 44 0 191 586 0227 sales rigelmedical com calibration seaward co uk www seaward co uk www rigelmedical com Prior to returning your unit for service please contact our service department to obta...

Страница 71: ...monitor setup Note that some monitor types might not be able to display low range sats Heart rate 30 300 bpm Accuracy 1 bpm Compatibility Beijing Choice Criticare GE Tuffsat Masimo Mindray Nellcor Nel...

Страница 72: ...settings Auto Mode View Data Data Transfer Download to PC Upload fro PC Configuration Data Load T nT Logo Setup Test Sequences Test Codes Asset Trace Variables System Config Bluetooth Favourites User...

Отзывы: