43 |
Page Rigel Medical
SP-SIM
user Manual V2.2
Run Mode
Semi-Automatic allows the user the enter data after each individual test that
forms a test sequence.
Full Automatic will run through the test sequence and provides an opportunity to
enter data at the end of the complete sequence.
Test Period
The re-test period allows the user to include the next test date on printed labels and PC
downloads that make it easier to schedule future work. More details on printing labels
see 6.2)
Trace Variables
To increase the traceability of the safety test, the user can include valuable data with
the test record such as Re-test period and Asset variables (see 4.2 for more
information). Trace variables can be added once the above data has been entered and
confirmed (See next chapters for more details
6.1. Performing an Automatic Test:
The automatic mode provides an option to test all available tests (SPO2 simulations,
Chronometer Test) in an automatic sequence.
6.1.1. Failure Menu
When a specific test fails, the user will be provided with a number of options depending
on the nature of the fault. The Failure Menu will enable the user to:
Restart Test (Individual test)
Restart Test Sequence (Whole Test
sequence)
Resume Test (Skip the failed test)
End Test Sequence (store result
and follow options. See 6.3)
Abort
Test
Sequence
(no
information stored, return to home
screen)
The following procedures describe the setup and completion of an automatic test
sequence. The steps are applicable to all available test sequences. As example we
describe a default test sequence.