46 |
Page Rigel Medical
SP-SIM
user Manual V2.2
Following the initial simulation at 98%, HR 80 bpm, the tester offers a result screen as
shown above. Enter the data from the monitor into the results screen and confirm. The
next test in this example is a chronometer test, measuring the time taken for the monitor
to establish a reading from 90% to 85% (target) SPO2;
After 6 seconds the SPO2 monitor displayed the required value at which the green
button is pressed to stop the chronometer. The results screen is then displayed
to confirm the reading as shown below;
Following this test in this example, the SP-SIM is setup to perform an an alarm check as
shown below;