5010e_513.doc / 22.01.10
ERROR MESSAGE / CORRECTION
9
ERROR MESSAGE / CORRECTION
9.1
GENERAL NOTE
Faulty input (e.g. wrong method number or wrong factor), recognized by the user, can be corrected by filling up
the respective entry field with any signs. After replenishing beyond the last position the faulty input is deleted
and the entry field is free again for the renewed correct input.
Error messages by the device are carried out either exclusively via a signal tone (chapter 9.2 - ACOUSTIC
ERROR MESSAGES) or as combination signal tone and display.
In the display errors are shown as plaintext (chapter 9.3 - PLAINTEXT ERROR MESSAGES)
… or coded with an error number (chapter 9.4 - CODED ERROR MESSAGES).
Each error message has always to be confirmed with [E].
9.2
ACOUSTIC ERROR MESSAGES
When pressing a key which is not permitted or not meaningful a deeper signal tone still sounds as error
message after the higher signal tone (which is to confirm the keystroke, can be switched off according to
chapter 7.2.14 - Key signal ON / OFF). In the display no corresponding error message appears parallel to this.
The operation of the device can directly be continued by the correct keyboard entry.
9.3
PLAINTEXT ERROR MESSAGES
RANGE MIN.
The programmed low limit was under-run by the measurement.
RANGE MAX.
The programmed upper limit was exceeded by the measurement.
RANGE R^2
The square of the correlation coefficient R lies at the kinetic measuring below the
programmed low limit.
RANGE +/-
At the kinetic measuring the procedure of the kinetic is wrong (increasing / falling).
NO METHOD
Dialed method is not programmed. Select other method according to method list.
HEATING OFF
Heating / cooling is off during temperature calibration.
TEMP. UNSTABLE
Temperature is unstable during temperature calibration.
9.4
CODED ERROR MESSAGES
No. (possible)
Causes
Remedy
1
method is write protected, method cannot be cleared
by special software
2
check sum of a freely programmed method is wrong
program new method
3
forbidden input, wrong number format
repeat input in permitted area
4
method not available
method editor:
check method no.
5
dark value is absolutely too high (> 16 bit) or higher as the
measurement,
ADC overflow
repeat dark adjustment;
check lamp / filter; check blank
6
all multiplexer positions are too bright/too dark at setting to
zero
repeat dark adjustment;
check filter / lamp; check blank
7
mathematical overflow, at measurement calculation
check filter; check standard;
check measuring solution