5010e_513.doc / 22.01.10
5.4.2
Calculation procedure 2 (C/F/Rb)
Method at which the difference of sample value A
S
and reagent blank A
RB
is multiplied with a given factor F.
The reagent blank A
RB
is entered or measured once.
Calculation procedure .......................................................CP 2
Characteristic ............................................................ C / F / Rb
Method ....................................................End Point with Factor
Calculation formula .......................................C = F
∗
(A
S
- A
RB
)
Factor............................................................... given / entering
Reagent blank ........................................entering or measuring
PHOTOMETER 5010 #7000
V5.13a 02/09/09 D
LAB.: RIELE GMBH+CO KG
DATE: 09/08/09
TIME: 08:52:24
METHOD 21: HDL-C
PROGRAM: 2
FACTOR: 325
WAVELENGTH: 546nm
TEMPERATURE: 37C
MEAS. VOLUME: 900ul
WASH VOLUME: 1000ul
DELAY: 5s
UNIT: mg/dl
- - - - - - - - - - - -
MEASURE BLANK
Rb[A]: 0.058
NO. ABS. RESULT
- - - - - - - - - - - -
1 1.064 327
- - - - - - - - - - - -
2 1.188 367
- - - - - - - - - - - -
3 1.340 417
- - - - - - - - - - - -
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Æ
Insert / measure zero solution
Æ
Insert / measure reagent blank
Æ
Insert / measure sample
Æ
Insert / measure sample
Æ
Insert / measure sample
ROBERT RIELE GmbH & Co KG
Page 25
5 - CALCULATION PROCEDURES