5010e_513.doc / 22.01.10
5.4.3
Calculation procedure 3 (C/F/Sb)
Method at which the difference of sample value A
S
and sample blank A
SB
regarding the amount is multiplied with
a given factor F. The sample blank A
SB
is measured before every test.
Calculation procedure .......................................................CP 3
Characteristic ............................................................ C / F / Sb
Method ....................................................End Point with Factor
Calculation formula ....................................... C = F
∗
|A
S
- A
SB
|
Factor............................................................... given / entering
PHOTOMETER 5010 #7000
V5.13a 02/09/09 D
LAB.: RIELE GMBH+CO KG
DATE: 09/08/09
TIME: 09:45:32
METHOD 23: BILIRUBIN
PROGRAM: 3
FACTOR: 12.80
WAVELENGTH: 546nm
TEMPERATURE: 37C
MEAS. VOLUME: 900ul
WASH VOLUME: 1000ul
DELAY: 5s
MAX.UNITS: 8.0
UNIT: mg/dl
- - - - - - - - - - - -
MEASURE BLANK
NO. ABS. RESULT
1 1.000 4.21
Sb[A]: 0.671
- - - - - - - - - - - -
2 1.215 4.25
Sb[A]: 0.884
- - - - - - - - - - - -
3 1.033 4.23
Sb[A]: 0.702
- - - - - - - - - - - -
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Æ
Insert / measure zero solution
Æ
Insert / measure sample blank
Æ
Insert / measure sample
Æ
Insert / measure sample blank
Æ
Insert / measure sample
Æ
Insert / measure sample blank
Æ
Insert / measure sample
ROBERT RIELE GmbH & Co KG
Page 26
5 - CALCULATION PROCEDURES