5010e_513.doc / 22.01.10
Printout of the monthly memory of a method with serum 2:
** MONTHLY MEMORY **S2 *
DATE: 09/08/09
PHOTOMETER 5010 #7000
V5.13a 02/09/09 D
LAB.: RIELE GMBH+CO KG
METHOD 25: GLUCOSE
UNIT: mmol/l
SERUM NO. 5
ID LT-SYS abnormal
LOT G312
COMPANY LABOR+TECH
DATE MAY 06
REQUIRED 14.4
MIN.VALUE 12.1
MAX.VALUE 16.7
QC VALUES n: 20
MEAN m: 13.860
STD.DEVIATION s: 1.012
COEFF.OF VAR CV: 7.298
02.15.06 13.54 -1s 1
02.14.06 14.07 +1s 1
02.13.06 14.69 +1s 1
02.12.06 13.50 -1s 3
02.11.06 14.68 +1s 3
02.10.06 15.33 +2s 1
02.09.06 15.99 +3s 1
02.08.06 15.38 +2s 2
02.07.06 14.61 +1s 1
02.06.06 13.70 -1s 1
02.05.06 12.74 -2s 1
02.04.06 12.13 -2s 1
02.03.06 12.65 -2s 2
02.02.06 13.11 -1s 1
02.01.06 13.88 +1s 3
01.31.06 13.51 -1s 3
01.30.06 13.24 -1s 3
01.29.06 12.50 -2s 1
01.28.06 12.74 -2s 2
01.27.06 13.30 -1s 1
Output dialog
If at least 20 readings are stored in the
monthly memory, these are indicated in the
Levey Jennings plot. In this representation the
deviations can be controlled visually and thus
tendencies or systematic errors be better
recognized.
The keys for the printout are located next to
the curve diagram:
[PRN-N] Start the normal printout of the data
of the current monthly memory. The single
data of the readings are not printed thereby.
[PRN-E] Start the extended printout of the
data of the current monthly memory. As
shown in the example left, also the single data
of the readings are printed.
[EXIT] Return to QC menu
P R N - N
P R N - E
E X I T
ROBERT RIELE GmbH & Co KG
Page 54
7- UTILITY PROGRAMS
Y PROGRAMS