5010e_513.doc / 22.01.10
7.2.4.4 Media test
With the media test the bubble detector can
be checked and adjusted.
In the second line to the left of the medium
WATER the current sensitivity level of the
bubble detector is shown.
Sip distilled water by [Sipping lever P]. Be
sure that the aspiration tube is completely
filled with water.
Enter new sensitivity level step by step by
numeric key [0] up to [7], starting with 0, until
the display changes from “AIR” to “WATER”.
Then increase the sensitivity level by one
further step.
Empty the aspiration tube by [Sipping lever P].
The display changes from “WATER” to “AIR”.
The adjustment is permanently saved by [E].
M E D I A T E S T
4 W A T E R 2 4 5
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
ROBERT RIELE GmbH & Co KG
Page 49
7- UTILITY PROGRAMS