5.4.9
Calculation procedure 9 (FTK/F/Rb)
Method at which a reagent blank is measured after an incubation time (
⇒
A
RB,0
) and after a reaction time (
⇒
A
RB,1
) and also a sample after an incubation time (
⇒
A
S,0
) and after a reaction time (
⇒
A
S,1
).
The difference from the change of the test and the change of the reagent blank is multiplied by a predefined
factor F. The reagent blank A
RB
is entered or measured once.
During the procedure the dialog asks for the use of a reagent blank. The default value is OFF. To continue with-
out reagent blank press [ENTER].
After each measurement the next sample can be measured with [NEXT].
With [RESULT] it is possible to measure the same sample again.
Calculation procedure .......................................................CP 9
Characteristic ........................................................ FTK / F / Rb
Method ................................................. Fixed Time with Factor
Calculation formula ......... C = F
∗
( |A
S,0
- A
S,1
| - |A
RB,0
- A
RB,1
| )
Factor............................................................... given / entering
Reagent blank ........................................entering or measuring
PHOTOMETER 5010 #7000
V5.13a 02/09/09 D
LAB.: RIELE GMBH+CO KG
DATE: 09/08/09
TIME: 09:47:33
METHOD 29: CK-MB
PROGRAM: 9
FACTOR: 2751.3
WAVELENGTH: 340nm
TEMPERATURE: 37C
MEAS. VOLUME: 900ul
WASH VOLUME: 1000ul
INCUBATION: 120s
REACTION: 180s
MAX.UNITS: 1500
UNIT: U/l
- - - - - - - - - - - -
MEASURE BLANK
Rb[A]: 0.000
NO. ABS. RESULT
- - - - - - - - - - - -
1 1.005 910.7
DELTA [A]: 0.331
- - - - - - - - - - - -
2 1.029 1128.1
DELTA [A]: 0.410
- - - - - - - - - - - -
3 0.829 1381.2
DELTA [A]: 0.502
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Æ
Insert / measure zero solution
Without reagent blank (insert / measure optionally)
Æ
Insert / measure sample
Æ
Insert / measure sample
Æ
Insert / measure sample
ROBERT RIELE GmbH & Co KG
Page 32
5 - CALCULATION PROCEDURES