7.2.2 Multi-standard
functions
7.2.2 Multi-standard
functions
Before curve data of a method with multi-standard can be processed, the method must have been established
in the method editor (chapter 6 - METHOD EDITOR). A curve without corresponding method cannot be
Before curve data of a method with multi-standard can be processed, the method must have been established
in the method editor (chapter 6 - METHOD EDITOR). A curve without corresponding method cannot be
processed! The term “curve no” has the same meaning as “method no”.
If using a multi-standard method later on, pay attention that all extinction values of the samples lie within the
range of the curve bases. Values outside of the extinction range cannot be calculated. In this case “+-“ is shown
and “<<< >>>” printed instead of the reading.
Fig. 7.2.2.1
Main window of multi-standard functions
[MEASURE CURVE] After query of a curve
number and the first standard the program
changes automatically to the method selection
window. There the given parameters of the
respective method can be again controlled
and/or changed. All further standards are
queried during the following procedure.
The program for measuring the multi-
standards branches out automatically to the
calculation procedures as follows:
CP1
→
CP5
CP2
→
CP6
CP3
→
CP7
CP4
→
CP8
CP9
→
CP10
CP11
→
CP12
CP14
→
CP14
Print-out of curve:
For measuring in another standard
method than suggested, start measuring
multi-standards in each other selected
standard method (before measuring the first
standard) by [MODE]
[
MODE] [M-STD]. Curve
number and standards are queried
accordingly.
[MEASURE CURVE] After inquiry of a curve
number start of measuring of standard directly
with a basic method which is based on
standard. This special mode ends automati-
cally after measuring or after selection of a
factor method. The standard of the selected
basic method represents the first standard
which can be measured. All further standards
are queried during the measuring execution in
the special mode.
[PRINT CURVE] After inquiry of the curve
number print-out on the internal printer or via
the serial interface.
For measuring in a multi-standard
method at least 2 bases with A and C must be
defined!